Stereo System User Manual
Chapter 4  95
Option P03 - Preamplifier
Other Preamp Specifications
Description Specifications Supplemental Information
Displayed Average Noise Level 
(DANL) − Preamp On
(Option P03)
a
a. DANL for zero span and swept is normalized in two ways and for two reasons. DANL is measured in a 
1 kHz RBW and normalized to the narrowest available RBW, because the noise figure does not depend 
on RBW and 1 kHz measurements are faster. The second normalization is that DANL is measured with 
10 dB input attenuation and normalized to the 0 dB input attenuation case, because that makes DANL 
and third order intermodulation test conditions congruent, allowing accurate dynamic range estimation 
for the analyzer.
Input terminated,  
Sample or Average detector 
Averaging type = Log 
0 dB input attenuation 
IF Gain = Any setting
Refer to the footnote for 
 Band Overlaps on page 15.
1 Hz Resolution Bandwidth
Preamp On
20 to 30 °C 5 to 50 °CTypical  Nominal 
Option P03
 100 kHz to 1 MHz
b
b. Specifications apply only when the Phase Noise Optimization control is set to “Best Phase Noise at off-
set > 30 kHz.”
−146 dBm
 1 MHz to 10 MHz −161 dBm
 10 MHz to 2.1 GHz −161 dBm −159 dBm −163 dBm
 2.1 GHz to 3.6 GHz −160 dBm −158 dBm −162 dBm