User`s guide
Key Reference 4
N9320A Spectrum Analyzer 99
For more details on how to measure the phase noise of a 
signal, refer to “Measuring Phase Noise” on page 55.
Marker Trace
Auto 1 2 3 4
Places a marker on the selected trace, if no markers have 
been previously turned on. If a marker is currently active, 
press  Marker Trace  until  1,  2,  3, or 4  is underlined: the active 
marker moves to the selected trace.
Selecting  Auto  mode moves the marker to the automatically 
selected trace. The trace selection will search for the lowest 
numbered trace in the following mode order: Clear-write, 
Max-hold, Min-hold and View. If there are no traces in any of 
these modes, it selects trace 1.
Key access: 
 > More
Marker Table Compresses the graticule and displays marker information in 
a table. The information includes the marker number, trace 
number, marker type, X axis value, and the amplitude.
Key access: 
 > More
Optimize 
Phase Noise
This softkey is valid only when span fall in the 
range of 100 Hz and 100 kHz. After enabling 
this softkey, the 1st LO Loop filter bandwidth 
will be changed from approx. 5 kHz to 120- 150 
kHz automatically, which optimizes the phase 
noise close in to the carrier.
Key access: 
 > Function > Phase Noise
Marker
NOTE
Optimize Phase noise (On) will worsen phase noise accuracy far away 
from the carrier. When testing phase noise far away from the carrier, 
select Optimize Phase noise (Off).
Marker
On
Off
Activates/deactivates the marker table.
Key access: 
 > More > Maker Table
Prev Page
Next Page
displays the previous/next page of the marker 
table.
Key access: 
 > More > Maker Table
Marker
Marker
Marker