User`s guide
110 N9320A Spectrum Analyzer
4 Key Reference
Preset/System
Accesses the Preset and System softkeys. When the analyzer 
is in the remote mode, pressing   returns the analyzer 
to the local controlmode and enables front- panel inputs.
Preset Provides a known state of the instrument for making 
measurements. The analyzer allows you to select whether a 
user preset, or factory preset is performed.
If Preset Type  is set to Factory, pressing   > Preset  performs 
a factory preset, which performs the following:
• Resets the analyzer to Spectrum Analyzer (SA) mode
• Brings up the Frequency  menu
• Sets certain conditions to their default values
• Clears both the input and output buffers and clears all 
trace data
• Sets the amplitude values of trace 2, 3 and 4 to the 
bottom of the screen
• Turn amplitude- correction factors off, but remain in 
analyzer memory
• Turn limit line testing off, but the limit line tables remain 
in analyzer memory
See Table Factory Preset States as follows for the conditions 
established by performing a factory preset.
When  Preset Type  is set to User, pressing   > Preset 
displays the softkeys User Preset and Factory Preset, you then 
select the specific preset as you need.