View/Trace
Accesses the trace keys to store and manipulate trace 
information. Each trace consists of a series of points in 
which amplitude data is stored. The analyzer updates the 
information for any active trace with each sweep.
Select Trace
1 2 3 4
Selects the trace for current use.
Key access: 
Clear Write Erases any data previously stored in the selected trace, then 
continuously displays any signals during each sweep of the 
analyzer. This function is activated for trace 1 at power- on 
and  Factory Preset.
Key access: 
Max Hold Retains the maximum level for each trace point of the 
selected trace. Updates the trace point value if a new 
maximum level is detected in successive sweeps.
Key access: 
Min Hold Retains the minimum level for each trace point of the 
selected trace. Updates each minimum trace point if a new 
minimum level is detected in following sweeps.
Key access: 
View Holds and displays the amplitude data of the selected trace. 
The trace register is not updated as the analyzer sweeps. 
press  View to retrieve a stored trace previously deactivated 
by pressing Blank.
Key access: 
Blank Stores the amplitude data for the selected trace, then 
removes the trace from the display. The selected trace 
register no longer updates as the analyzer sweeps. To 
activate this function for traces 2, 3 and 4 at power- on press 
Preset (when Preset Type is set to Factory).
Key access: 