User`s guide
124 N9320A Spectrum Analyzer
5 One-Button Measurements
Occupied BW
Occupied Bandwidth integrates the power of the displayed 
spectrum. The power- bandwidth routine first computes all 
signal power contained in the trace, then calculates the span 
containing the power of the specified percentage. The 
percentage value default is 99% of the occupied bandwidth 
power.
For 99% occupied power bandwidth, Markers are placed at 
the frequencies on either side of 99% of the power. 1% of the 
power is evenly distributed outside the markers. The 
difference between the marker frequencies is the 99% power 
bandwidth.
This is the main menu of occupied bandwidth measurement 
including all related parameters and measurement results.
Figure 29 Occupied Bandwidth Measurement
The occupied bandwidth function also indicates the 
difference between the center frequency and the center 
frequency of the channel. This “Transmit Freq Error”
f1 f2
Measurement
Result