User`s guide
One-Button Measurements 5
N9320A Spectrum Analyzer 125
is defined as the difference between the (f
1 + f2)/2 and the 
tuned center frequency of the signal.
Avg Number
On Off
Select  On to specify the number of measurement averages 
used when calculating the measurement result. The average 
is displayed at the end of each sweep. Select Off to disable 
measurement averaging. The range is from 1 to 1000.
Key access: 
 > Occupied BW
Avg Mode
Exp Repeat
Toggles the termination control type used for the averaging 
function
  between  Exp  and  Repeat. This  determines  the 
averaging action after the specified number of measurements 
(average count) is reached. 
When you select Exp, each successive data acquisition after 
the average count is reached is exponentially weighted and 
combined with the existing average. Exponential averaging 
weights new data more than old data, which facilitates 
tracking of slow- changing signals. The average is displayed at 
the end of each sweep.
When you select Repeat, after reaching the average count, all 
previous data is cleared and the average count is reset to 1 
and the averaging process is restarted.
Key access: 
 > Occupied BW
Max Hold
On Off
Toggles the maximum hold trace feature between On  and  Off. 
Maximum hold displays and holds the maximum responses 
of a signal.
Key access: 
 > Occupied BW
Occ BW% Pwr Edits the percentage of signal power used when determining 
the occupied bandwidth.
Key access: 
 > Occupied BW
NOTE
To get the correct measurement result, first set the center frequency, 
reference level, and channel spacing.
Meas
Meas
Meas
Meas