User`s guide
126 N9320A Spectrum Analyzer
5 One-Button Measurements
OBW Span Specifies the integration range used to calculate the total 
power from which the percent occupied bandwidth is then 
calculated. The analyzer span will be set to the same value 
as the OBW Span for the measurement. OBW Span should 
be set to an approximate multiple of 2
  times  the  expected 
occupied
  bandwidth  result.  The  range  is  from  100  Hz  to  3 
GHz.
Key access: 
 > Occupied BW
x dB Specifies the power level used to determine the emission 
bandwidth as the number of dB down from the highest 
signal point, within the occupied bandwidth span. This 
function is an independent calculation from the OBW 
calculation. The x dB Bandwidth result is also called the 
emissions bandwidth, or EBW. Two arrows will be set to the 
specified dB value below the maximum power level and the 
total power between the arrows will be computed. 
Frequencies f1 and f2 are determined as the furthest 
frequencies x dB below and above maximum power level, 
respectively. The emission bandwidth is then calculated as f2 
– f1 as shown by the arrows indicated in Figure 29  (note 
that the diamond markers are for the OBW function). The 
range is from –100.0 to –0.1 dB.
Key access: 
 > Occupied BW
Meas
Meas
NOTE
The asterisk next to the x dB bandwidth value in Figure 29 indicates 
the results may not have been determined with optimal analyzer 
settings. If emission bandwidth is your primary interest, select Max 
Hold. Then change detector mode to peak by pressing 
 > 
Detector > Positive Peak. Acquiring peak data ensures accuracy of 
this measurement.
Det/
Display