User`s guide
One-Button Measurements 5
N9320A Spectrum Analyzer 129
Meas Type Specifies the reference for the measurement, either Tota l Pw r 
Ref  or  PSD Ref. Relative values are displayed referenced to 
either the total power (Tota l  P w r R ef ) or the power spectral 
density (PSD Ref) measured in the main channel.
Key access: 
 > ACP
Method
IBW RBW
Enables you to select the measurement method. The 
resolution bandwidth (RBW) method is most useful for 
measuring cdmaOne signals; the integration bandwidth 
method (IBW) is preferred for other signals.
Key access: 
 > ACP > More
Tota l P w r R ef
(PSD Ref)
Auto Man
Sets the adjacent channel power reference to automatic or 
manual when Meas Type  is set to Tot al P wr R ef . When select 
Auto, the carrier power result reflects the measured power 
value in the carrier. When set to Man, the last measured 
value is captured and held, or may be entered by the user. 
Relative values are displayed, referenced to the total power 
measured in the main channel.
When  Meas Type is set to PSD Ref, pressing this key sets the 
power spectral density in the carrier (main channel) that is 
used to compute the relative power spectral density values 
for the offsets.When the PSD Ref state is set to Auto, this 
will be set to the measured carrier power spectral density.
Key access: 
 > ACP > More
Limits
On Off
Turns limits testing on or off. When select On, each offset is 
compared to its upper and lower offset limit. In those cases 
where the power exceeds the limit, the calculated result will 
be displayed in red to indicate a failure. If there are any 
failures, the pass/fail indicator area show “FAIL” in red; if 
there are none, it shows “PASS” in green when using IBW 
method. Neither “FAIL” nor “PASS” will be displayed when 
using RBW method. Any offsets that are in the off state are 
not measured and no results will be displayed.
Key access: 
 > ACP > More
Meas
Meas
Meas
Meas