130 N9320A Spectrum Analyzer
5 One-Button Measurements
Intermod (TOI)
The third order intermodulation (TOI) measurement computes 
and displays the output intercept point (IP3), and places 
arrows upon the trace to indicate the measured signals and 
third- order products.
Figure 31 TOI Measurement
Avg Number
On Off
Set  Avg Number  to  On to specify the number of measurement 
averages used when calculating the measurement result. The 
average will be displayed at the end of each sweep. Setting 
Avg Number  to  Off disables measurement averaging.
Key access: 
 > More> Intermod (TOI)
Avg Mode
Exp Repeat
Selects
 the type  of termination control used for the averaging 
function (Exp/Repeat). This determines the averaging action 
after the specified number of measurements (average count) 
is reached.
When  Avg Mode  is  Exp, Each successive data acquisition
  after 
the  average  count  is  reached  is exponentially weighted and 
combined with the existing average. Exponential averaging 
weights new data more than old data, which facilitates 
tracking of slow- changing signals. The average is displayed at 
the end of each sweep.
Meas