User`s guide
One-Button Measurements 5
N9320A Spectrum Analyzer 135
a. Abs Stop is set to Couple and Fail Mask is set to Abs AND Rel.
a.  Res BW  is set to 100 kHz and Fail Mask is set to Rel.
Table 2 Offsets & Limits Defaulted for W-CDMA (3GPP) Mobile Station Measurements
a
Offset Start 
Freq
(MHz)
Stop 
Freq
(MHz)
Res BW
(kHz)
Abs 
Start
(dBm)
Abs 
Stop
(dBm)
Rel 
Couple
Rel 
Start
(dBc)
Rel Stop
(dBc)
Meas 
BW
A. On 2.515 3.485 30.00 -71.07 -71.07 Man -35.23 -49.78 1
B. On 4.000 7.500 1000.0 -55.84 -55.84 Man -35.50 -39.00 1
C. On 7.500 8.500 1000.00 -55.84 -55.84 Man -39.00 -49.00 1
D. On 8.500 12.000 1000.00 -55.84 -55.84 Couple -49.00 -49.00 1
E. Off 12.500 15.000 1000.00 -55.84 -55.84 Couple -49.00 -49.00 1
Table 3 Offsets & Limits Defaulted for 802.11a and HiperLAN/2 Radio Standards
a
Offset Start 
Freq
(MHz)
Stop 
Freq
(MHz)
Abs 
Start
(dBm)
Abs 
Stop
(dBm)
Abs 
Couple
Rel 
Start
(dBc)
Rel Stop
(dBc)
Rel 
Couple
Meas 
BW
A. On 9.00 11.00 16.00 -4.00 Man 0.00 -20.00 Man 1
B. On 11.00 20.00 -4.00 -12.00 Man -20.00 -28.00 Man 1
C. On 20.00 30.00 -12.00 -24.00 Man -28.00 -40.00 Man 1
D. On 30.00 50.00 -24.00 -24.00 Couple -40.00 -40.00 Couple 1
E. Off 50.00 100.00 -24.00 -24.00 Couple -40.00 -40.00 Couple 1