User`s guide
One-Button Measurements 5
N9320A Spectrum Analyzer 137
Res BW
Auto Man
Specifies the resolution bandwidth used in measuring the 
offset pair. When set to Auto, the Res BW from the default 
tables for “Offset/Limits” on page 134  are used. When set to 
Man, the allowed range is 1 kHz to 1MHz.
Key access: 
 > More > Spectrum Emission Mask > Offset/Limits
Abs Start To enter an absolute level limit at Start Freq  for the currently 
selected offset ranging from –200.00 to +50.00 dBm with 4 
significant digits.
Key access: 
 > More > Spectrum Emission Mask > Offset/Limits 
> More
Abs Stop
Couple Man
To enter an absolute level limit at Stop Freq  for the currently 
selected offset ranging from –200.00 to +50.00 dBm with 4 
significant digits, and to toggle this function between Couple 
and  Man. If set to Couple, this is coupled to Abs Start  to make 
a flat limit line. If set to Man,  Abs Start  and  Abs Stop  the user 
can enter different values to make a sloped limit line.
Key access: 
 > More > Spectrum Emission Mask > Offset/Limits 
> More
Rel Start Enters a relative level limit at Start Freq  ranging from –150.00 
to +50.00 dBc with 4 significant digits.
Key access: 
 > More > Spectrum Emission Mask > Offset/Limits 
> More
Rel Stop
Couple Man
Enters a relative level limit at Stop Freq  ranging from –150.00 
to +50.00 dBc with 4 significant digits, and to toggle this 
function between Couple  and  Man. If set to Couple, this is 
coupled to Rel Start  to make a flat limit line. If set to Man,  Rel 
Start  and  Rel Stop  the user can enter different values to make 
a sloped limit line.
Key access: 
 > More > Spectrum Emission Mask > Offset/Limits 
> More
Fail Mask Displays the menu to select one of the following logic keys 
for fail conditions between the measurement results and the 
test limits: Absolute,  Relative,  Abs AND Rel, and Abs OR Rel.
Key access: 
 > More> Spectrum Emission Mask > Offset/Limits 
> More > More
Meas
Meas
Meas
Meas
Meas
Meas