User`s guide
138 N9320A Spectrum Analyzer
5 One-Button Measurements
Absolute Sets the limit test to show Fail if one offset is 
larger than the corresponding limit for Abs Start 
and/or  Abs Stop. This is the default selection for 
each offset for BTS measurements.
Key access: 
 > More> Spectrum Emission Mask 
> Offset/Limits > More > More > Fail Mask
Relative Sets the limit test to show Fail if one offsets 
spectrum emission mask measurement results is 
larger than the corresponding limit for Rel Start 
and/or  Rel Stop.
Key access: 
 > More> Spectrum Emission Mask 
> Offset/Limits > More > More > Fail Mask
Abs AND Rel Set the limit test to show Fail if one offset’s 
spectrum emission mask measurement results is 
larger than both the corresponding limit for Abs 
Start  and  Abs Stop  and the limit for Rel Start  and 
Rel Stop.
Key access: 
 > More > Spectrum Emission Mask 
> Offset/Limits > More > More > Fail Mask
Abs OR Rel Set the limit test to show Fail if one offset’s the 
absolute spectrum emission mask measurement 
results is larger than the corresponding limit 
for Abs Start  and  Abs Stop  or the limit for Rel 
Start  and  Rel Stop.
Key access: 
 > More > Spectrum Emission Mask 
> Offset/Limits > More > More > Fail Mask
Meas BW Specifies the bandwidth to use when measuring the offset. 
When using the front panel, only the currently selected 
offset is affected. All the offsets are affected when remote 
control. This is the multiplier applied to the resolution 
bandwidth to determine the bandwidth to integrate when 
calculating the power. Meas BW is now restricted to be 1 
and not changeable.
Key access: 
 > More > Spectrum Emission Mask > Offset/Limits 
> More > More
Meas
Meas
Meas
Meas