Technical data
Command Reference 5
N9320A Programmer’s Guide 97
Set the Measurement Method
[:SENSe]:ACPower:METHod IBW|RBW
[:SENSe]:ACPower:METHod?
Enables you to set the measurement method. The resolution 
bandwidth method is most useful for measuring the signals 
the bandwidths of which is approximate to the RBW of the 
Analyzer; the integration bandwidth method is preferred for 
other signals.
*RST: IBW
Example: ACP:METH:RBW
Key access:
 > ACP > More > Method
Set the Ref BW
[:SENSe]:ACPower:OFFSet:LIST:BAND-
width|BWIDth[:INTegration] 
<freq>,<freq>,<freq>,<freq>,<freq>,<freq>
[:SENSe]:ACPower:OFFSet:LIST:BANDwidth|BWIDth[:I
NTegration]?
Sets the reference bandwidth (integration bandwidth) for 
each offset.
*RST: The default value is defined by the selected Offset label. 
Table 1 Offset parameters in ACP measurement
Meas
Offset State Frequency Integ. BW
A ON 3 MHz 2 MHz
B OFF 0 Hz 2 MHz
C OFF 0 Hz 2 MHz
D OFF 0 Hz 2 MHz
E OFF 0 Hz 2 MHz
F OFF 0 Hz 2 MHz