Technical data
120 N9320A Programmer’s Guide
5 Command Reference
Typ e  of Dete c tion
[:SENSe]:DETector[:FUNCtion]NEGative|POSi-
tive|SAMPle|AVERage|RMS|NORMAL
[:SENSe]:DETector[:FUNCtion]?
Specifies the detection mode. For each trace interval 
(bucket), average detection displays the average of all the 
samples within the interval. The averaging can be done using 
two methods:
the power method (RMS)
the video method (Y Axis Units)
The method is controlled by   > > Avg Type.
• Negative peak detection displays the lowest sample taken 
during the interval being displayed.
• Positive peak detection displays the highest sample taken 
during the interval being displayed.
• Sample detection displays the sample taken during the 
interval being displayed, and is used primarily to display 
noise or noise- like signals. In sample mode, the 
instantaneous signal value at the present display point is 
placed into memory. This detection should not be used to 
make the most accurate amplitude measurement of non 
noise- like signals.
• Average detection is used when measuring the average 
value of the amplitude across each trace interval (bucket). 
The averaging method used by the average detector is set 
to either video or power as appropriate when the average 
type is auto coupled.
• Normal detection selects the maximum and minimum video 
signal values alternately. When selecting Normal detection, 
“Norm” appears in the upper- left corner.
*PST: Positive
Key access:  >
 Detector
BW/
Avg
Det/
Display