Technical data
Command Reference 5
N9320A Programmer’s Guide 57
Display the Limit Line
:CALCulate:LLINe[1]|2:DISPlay OFF|ON|0|1
:CALCulate:LLINe[1]|2:DISPlay?
Controls the display of the current limit line.
*RST: Off
Key access:
 > Limits > Limit 1|2 > Limit On Off
Test the Data Against the Limit Line
:CALCulate:LLINe[1]|2:FAIL?
Queries the status of the limit line testing. Returns a 0 if the 
data passes, and returns a 1 if there is a failure. This query 
value is valid only if margin or limit test is On. Use the 
command  :CALCulate:LLINe[1]|2:STATe OFF|ON|0|1 
to activate limit line testing.
Key access:
 > Limits > Limit 1|2 > Test On Off
Set the Margin Size
:CALCulate:LLINe[1]|2:MARGin <rel_ampl>
:CALCulate:LLINe[1]|2:MARGin?
Allows you to define the amount of measurement margin 
that is added to the designated limit line.
*RST: Off
Remarks: The margin must be negative for upper limit lines, and 
positive for lower limits.
Key access:
 > Limits > Limit 1|2 > Margin On Off