Technical data
Command Reference 5
N9320A Programmer’s Guide 69
Set the Delta Marker 
:CALCulate:MARKer[n]:X:DELTa <param>
:CALCulate:MARKer[n]:X:DELTa?
Activates a pair of markers, where each marker can be 
independently positioned on the trace. The marker readout 
shows the difference between the two markers. 
Key access:
 > Delta > Delta On Off
Set the Reference Marker
:CALCulate:MARKer[n]:X:REFerence <param>
:CALCulate:MARKer[n]:X:REFerence?
Specifies a pair of reference markers. The marker readout 
shows the difference between the two markers.
*RST: AUTO
Key access:
 > Delta > Delta Pair Ref Delta
Span Markers Span X Value
:CALCulate:MARKer[n]:X:SPAN <param>
:CALCulate:MARKer[n]:X:SPAN?
Change the frequency span of the designated span- type 
marker pair to position the markers at the desired trace X 
values. The value is in the X- axis units (which is usually 
frequency or time). 
The query returns the current X value frequency span of the 
designated markers. If span markers are not selected, the 
query returns the latest marker reading as a span (always 
positive).
Key access:
 > Delta > Span Pair
Marker
Marker
Marker










