User`s guide
Key Reference 4
N9320A Spectrum Analyzer 101
with a tracking generator. For example, 
connect the cables and a thru line, in place 
of the device to be measured, between the 
tracking generator and the analyzer input. 
Notice that the frequency response is not 
perfectly flat, showing the response of the 
cables, as well as the flatness of both the 
tracking generator and the analyzer. Now 
press  Store Ref (1->3),  Normalize (On). Notice 
that the displayed trace is now flat, or 
normalized. The position of the normalized 
trace can now be moved to a different 
position on the display by changing the 
normalized reference position. This may be 
useful if the device to be tested has 
positive gain, such as an amplifier. Now 
replace the thru line with the device under 
test, and an accurate measurement of the 
gain or loss can be made.
Key access: 
 > Tracking Generator > More 
> Normalize
Norm Ref Lvl Sets the level (in dB) of the normalized 
reference.
Key access: 
 > Tracking Generator > More 
> Normalize
Norm Ref 
Posn
Offsets the displayed trace without affecting 
the instrument gain or attenuation settings. 
This allows the displayed trace to be moved 
without decreasing measurement accuracy. 
The normalized reference position is 
indicated with a (>) character on the left 
side of the display and a (<) character on 
the right side of the display.
Key access: 
 > Tracking Generator > More 
> Normalize