User`s guide
102 N9320A Spectrum Analyzer
4 Key Reference
Ref Trace
View Blank
Allows you to view or blank the reference 
trace on the display.
Key access: 
 > Tracking Generator > More 
> Normalize
Tracking 
Peak
Activates a routine that automatically adjusts fine 
frequency tracking to obtain the peak response of the 
tracking generator on the spectrum analyzer display. 
Tracking Peak is performed in the current resolution 
bandwidth.
Key access:   > Tracking Generator>More
Man Track 
adj
Allows you to adjust the frequency of the tracking 
generator oscillator manually using the step keys, knob, 
or numeric keypad. The tracking adjust is tuned to 
maximize the amplitude of the trace.
Key access:
   > Tracking Generator>More
MODE
MODE
MODE
NOTE
When in the Tracking Generator mode, resolution bandwidths < 1 kHz 