User`s guide
114 N9320A Spectrum Analyzer
4 Key Reference
Sweep/Trig
Accesses the following softkeys to provide selection of the 
sweep mode and trigger mode of the analyzer:
Sweep Time
Auto Man
Selects the length of time the analyzer takes to tune across 
the displayed frequency span (or, in zero span, the time the 
analyzer takes to sweep the full screen). Reducing the sweep 
time increases the sweep rate. Change the sweep time by 
using the arrow keys, the knob, or the numeric keypad.
In non- zero spans, when the sweep time is auto- coupled, the 
analyzer selects the optimum (shortest) sweep time for the 
current settings. This selection process is influenced by 
several factors:
• The max. tuning rate of the spectrum analyzer
• The selected resolution and video bandwidth filters
• The max. sample rate of the analog- to- digital converter 
• The number of trace points
• The amplitude scale (logarithmic or linear)
• The detector mode
You may select a sweep time less than the auto- coupled 
value; however this may generate measurement errors. The 
minimum sweep time is 50 ms.
In zero span, the minimum sweep time is determined by the 
maximum sample rate of the ADC and the number of sweep 
points. The minimum sweep time is 2 microseconds. The 
maximum sweep time is 4000 seconds.
Key access:   
Sweep
Single
Pressing  Sweep Single  puts the analyzer in single- sweep 
mode. Press Sweep Single to initiate a sweep when in 
single- sweep mode.
Key access:   
Sweep Cont causes one sweep to follows another as soon as it is 
triggered. Pressing Preset  (When  Preset Type  is set to Factory) 
sets the sweep mode to continuous sweep.
Key access:   
Sweep/
Tri g
Sweep/
Tri g
Sweep/
Tri g