User`s guide
One-Button Measurements 5
N9320A Spectrum Analyzer 119
Avg Number
On Off
Specifies the number of measurement averages used when 
calculating the measurement result with Avg Number  (On). The 
range of the number is from 1 to 1000. Use the knob, the 
arrow keys or the numeric keypad to set the number. The 
average will be displayed at the end of each sweep. Selecting 
Off  disables measurement averaging.
Key access: 
 > Channel Power
Avg Mode
Exp Repeat
Toggles the average mode between Exp  or  Repeat. This 
determines the averaging action after the specified number 
of measurements (average count) is reached. 
When you select Exp, each successive data acquisition after 
the average count is reached is exponentially weighted and 
combined with the existing average. Exponential averaging 
weights new data more than old data, which facilitates 
tracking of slow- changing signals. The average will be 
displayed at the end of each sweep.
If you select Repeat, after reaching the average count, all 
previous data is cleared and the average count is back to 1.
Key access: 
 > Channel Power
Integ BW Specifies the range of integration used in calculating the 
power in a channel. For example, set the main (center) 
channel bandwidth. Note that the white frame indicates the 
integration bandwidth. Be sure the span  is set between 1 
and 10 times the integration bandwidth. Change integration 
bandwidth by using the knob and the arrow keys.
Key access: 
 > Channel Power
Chan Pwr Span Sets the analyzer span for the channel power measurement 
by using the knob and the arrow keys. Be sure the span  is 
set between 1 and 10 times the integration bandwidth.
Key access: 
 > Channel Power
Meas
Meas
Meas
Meas