User`s guide
120 N9320A Spectrum Analyzer
5 One-Button Measurements
Occupied BW
Occupied Bandwidth integrates the power of the displayed 
spectrum. The power- bandwidth routine first computes the 
power of all signal contained in the trace, then calculates 
the span which contains the power of a selected percentage. 
The percentage value defaults to 99% of the occupied 
bandwidth power.
For 99% occupied power bandwidth, Markers are placed at 
the frequencies on either side of 99% of the power. 1% of the 
power is evenly distributed outside the markers. The 
difference between the marker frequencies is the 99% power 
bandwidth.
This is the main menu of occupied bandwidth measurement 
including all related parameters and measurement results.
Figure 29 Occupied Bandwidth Measurement
The occupied bandwidth function also indicates the 
difference between the analyzer center frequency and the 
center frequency of the channel. This “Transmit Freq Error”
f1 f2
Measurement
Result