User`s guide
128 N9320A Spectrum Analyzer
5 One-Button Measurements
Spectrum Emission Mask
Spectrum Emission Mask measurement includes the in- band 
and out- of- band spurious emissions. As it applies to 
W- CDMA (3GPP), this is the power contained in a specified 
frequency bandwidth at certain offsets relative to the total 
carrier power. It may also be expressed as a ratio of power 
spectral densities between the carrier and the specified 
offset frequency band. For WLAN, the reference power is 
taken as the peak PSD in the signal, and all offset results 
are also the peak PSD in that offset as opposed to the 
integrated power.
Figure 32 Spectrum Emission Mask
Avg Number
On Off
Set  Avg Number  to  On to specify the number of measurement 
averages used when calculating the measurement result. The 
average will be displayed at the end of each sweep. Setting 
Avg Number  to  Off disables measurement averaging.
Key access: 
 > More > Spectrum Emission Mask
Meas Type Displays a menu where you can select a measurement 
reference type, Tota l P w r R ef   or  PSD Ref.
Key access: 
 > More > Spectrum Emission Mask
Meas
Meas