User`s guide
48 N9320A Spectrum Analyzer
3 Functions and Measurements
The signal produces harmonic distortion products (spaced 
200 MHz from the original 200 MHz signal) in the analyzer 
input mixer as shown in Figure 16.
4 Change the center frequency to the value of the first 
harmonic:
Press   > Next Peak
Press 
 > Mkr -> CF
5 Change the span to 50 MHz and re- center the signal:
Press   > 50 > MHz
Press   
Press 
 > Mkr -> CF
6 Set the attenuation to 0 dB:
Press   > Attenuation > 0 > dB
7 To determine whether the analyzer generates harmonic 
distortion products, first display the trace data in trace 2 as 
follows:
Press   > Select Trace > Trace 2
Press  Clear Write
8 Allow trace 2 to update (minimum two sweeps), then store 
the data from trace 2 and place a delta marker on the 
harmonic of trace 2:
Press   > View
Press   
Press 
 > Delta > Delta (On)
The analyzer display shows the stored data in trace 2 and 
the measured data in trace 1. The MarkerΔ indicator reads 
the difference in amplitude between the reference and active 
markers.
9 Increase the RF attenuation to 10 dB:
Press   > Attenuation > 10 > dB
Peak
Search
Marker
SPAN
Peak
Search
Marker
Amplitude
View/
Tra c e
View/
Tra c e
Peak
Search
Marker
Amplitude