User`s guide
Functions and Measurements 3
N9320A Spectrum Analyzer 51
2 Set one source (signal generator) to 300 MHz and the other 
source to 301 MHz, for a frequency separation of 1 MHz. Set 
the sources equal in amplitude as measured by the analyzer 
(in this example, they are set to –5 dBm).
3 Set the analyzer center frequency and span:
Press   > Preset (With  Preset Type of Factory)
Press   > Center Freq > 300.5 > MHz
Press   > Span > 5 > MHz
4 Reduce the RBW until the distortion products are visible:
Press and 
5 Move the signal to the reference level:
Press   
Press 
 > Mkr -> Ref Lvl
6 Reduce the RBW until the distortion products are visible:
Press and 
7 Activate the second marker and place it on the peak of the 
distortion product (beside the test signal) using the Next Pea:
Press 
 > Delta > Delta (On)
Press   > Next Peak
8 Measure the other distortion product:
Press 
 > Normal
Press   > Next Peak
NOTE
The coupler should have a high degree of isolation between the two 
input ports so the sources do not intermodulate.