5 Increase measurement sensitivity and smooth the noise:
Press    > Res Bw > 30 > KHz
Press    > Video Bw > 300 > KHz
A decrease in displayed amplitude is caused by tracking 
error
.
6 Use peak tracking to correct the frequency offset:
Press    > Tracking Generator > More > Tracking Peak
Tracking error occurs when the output frequency of the 
tracking generator is not exactly matched to the input 
frequency of the analyzer. The amplitude should return to 
the value that was displayed prior to the decrease in 
resolution bandwidth.
7 Connect the cable from the tracking generator output to the 
analyzer input. Store the frequency response in trace 3 and 
normalize:
Press    > Tracking Generator > More > Normalize > Store Ref (1
→ 3) > Normalize (On)
8 Reconnect the DUT to the analyzer and change the 
normalized reference position:
Press    > Tracking Generator > More > Normalize > Norm Ref 
Posn > 8
NOTE
To reduce ripples caused by source return loss, use 6 dB or greater 
output attenuation. Tracking generator output attenuation is normally 
a function of the source power selected. However, the output 
attenuation may be controlled in the Tracking Generator menu.