User`s guide
Key Reference 4
N9320A Spectrum Analyzer 77
When in continuous sweep mode, the specified number of 
averages is taken, then the averaging continues with each 
new sweep averaged in with a weight of  and the 
old average reduced by multiplying it by 
.
To turn averaging off, press Average (Off). 
Key access: 
Avg Type
Auto Man
Toggles the average type between automatic and manual.
Selecting Auto sets the analyzer to automatic averaging by the 
rules in Figure 2.
Figure 2 Auto Rules for Average Type
Selecting  Man sets the analyzer to manual averaging.
When trace average is on, the average type is shown on the 
left side of the display. Selecting an average type manually 
causes the analyzer to use that type without regard to other 
analyzer settings and sets Avg Type to Man. 
Key access: 
Video Avg Video averaging averages the data as appropriate 
  for the y- axis scale. When Avg Type > Video Avg  is 
  selected, “VAvg”  appears on the left side of the 
  analyzer display if Average  is  On.
  Key access:   > Avg Type
Pwr Avg 
  Power averaging is performed by converting the 
  trace data from logarithmic to linear power units, 
  and then averaging the power trace data. When 
  Avg Type > Pwr Avg  is selected, “PAvg”  appears on 
  the left side of the screen if Average  is  On.
  Key access:   > Avg Type
1
AverageNumber
------------------------------
AverageNumber
1
AverageNumber
------------------------------
–
BW/
Avg
Is Detector 
Average? 
T
yp
e=Power
T
yp
e=Video
Yes 
No 
BW/
Avg
BW/
Avg
BW/
Avg