Key Reference 5
N9340A User’s Guide 93
In FFT Mode
In FFT mode, the sweep time default is
auto- coupled. The submenu, sweep time, is invalid
in FFT mode.
Sweep
Key access: [BW/SWP] > {Sweep}
Activates the sweep mode function and accesses
the softkey submenus to select the trace sweep
functions.
Sweep
Toggles the analyzer between the continuous sweep
mode and the single sweep mode.
• Sweep (Single)
Puts the analyzer in a single sweep mode. In this
mode, pressing [BW/SWP] > {Single Sweep}
enables a sweep.
Key access: [BW/SWP] > {Sweep}
• Sweep (Cont)
Performs one sweep after another as soon as it
is triggered. Pressing [Preset] (When Preset Type
is set to Default) or turning the power on, sets
the sweep mode to continuous sweep.
Key access: [BW/SWP] > {Sweep}
Sweep Mode
An N9340A provides two sweep modes to meet
different sweep time requirements:
•Normal
Activates the default sweep mode.
Key access: [BW/SWP] > {Sweep} > {Sweep Mode}