Product specifications

17
In order to keep a communication system working reliably 
long term, you need to monitor the spectrum routinely. 
Regular spectrum monitoring is usually accomplished 
automatically by using remote control software to drive 
the spectrum analyzer. You can remotely control the HSA 
using SCPI commands over LAN and USB, or use the HSA 
PC software to monitor spectral changes. There are two 
modes that are used to set up the HSAs’ LAN IP address: 
static and DHCP.
In this demonstration, we will show how to set HSA’s IP 
address.
Instructions  for  HSA   Keystrokes for HSA
Set HSA’s IP address is in 
static mode
[Shift], {System}, {More 1 of 
2}, {Port Setting}, {IP Config}, 
{IP Address (Static)}
Set IP address such as: 
192.168.0. 12
[192.168.0.12], [Enter]
Set gateway such as: 
192.168.0.8
[192.168.0.8], [Enter]
Set subnet mask such as: 
255.255.255.0
[255.255.255.0], [Enter], 
{Apply}
Set HSA’s IP address 
automatically using DHCP 
mode
[Shift], {System}, {More 1 of 
2}, {Port Setting}, {IP Config}, 
{IP Address (DHCP)}, {Apply}
Demonstration 7
Remote control via LAN or USB port
The N934xC HSAs each have a total of six markers that 
can be used in normal or delta mode. In addition to the 
marker table and peak search, each N934xC HSA also has 
a peak table feature which automatically searches 10 max 
or min peak signals simultaneously and is updated after 
each sweep.
In this demonstration, we will use peak table to find the 
absolute values of FM signal sidebands. 
Note: To prevent data corruption when saving the table to a USB 
memory device, set the file directory to USB then reset to INT 
before removing the device.
Demonstration 8
Marker, marker table, and peak table
Instructions for the 
source 
Keystrokes for the 
source
Set 2 GHz center frequency, 
amplitude -10 dBm, 100 kHz 
deviation, 30 kHz rate, FM 
signal
[Preset], [FREQ], [2], {GHz}, 
[AMPTD], [-10], {dBm}, 
[FM/ФM], {FM On}, {FM Dev} 
[100], {kHz}, {FM Rate}, [30], 
{kHz}, [Mod On], [RF On]
Instructions  for  HSA  Keystrokes for HSA
Set center frequency to 2 GHz [Freq], [2], {GHz}
Set span to 500 kHz [Span], [500], {kHz}
Set reference level to -10 dBm [Amptd], {Ref Level}, {10}, 
{dBm}
Peak search [Shift], [Peak], {Peak Search}
Measure the absolute peak 
amplitude of FM sidebands 
using the peak search table. 
Peak search table will display 
up to 10 peaks (Figure 21)
{More 1 of 2}, {Peak Table}, 
{Peak Table On}
Change peak criterion [Shift], [Peak], {More 1 of 2}, 
{Peak Criterion}, user can set 
the peak threshold limit, peak 
excursion limit and peak type
Save peak table in a .csv 
format
[Shift], [Peak], {More 1 of 2}, 
{Peak Table}, {Export Table to 
CSV}, input file name, [Enter]
Turn the peak table off [Shift], [Peak], {More 1 of 2}, 
{Peak Table}, {Peak Table Off}
Figure 21. FM sidebands peak values in peak table