Technical data
Status Registers 3
Programmer’s Guide 29
The standard event status register contains following bits:
The standard event status register is used to determine the specific event 
that set bit 5 in the status byte register. To query the standard event 
status register, send the command *ESR?. The response will be the 
decimal sum of the bits which are enabled (set to 1). For example, if bit 
number 7 and bit number 3 are enabled, the decimal sum of the 2 bits is 
128 plus 8. So the decimal value 136 is returned. 
In addition to the standard event status register, the standard event status 
group also contains a standard event status enable register. This register 
lets you choose which bits in the standard event status register will set 
the summary bit (bit 5 of the status byte register) to 1. Send the *ESE 
<integer> command where <integer> is the sum of the decimal 
values of the bits you want to enable. For example, to enable bit 7 and 
bit 6 so that whenever either of those bits is set to 1, the standard event 
status summary bit of the status byte register will be set to 1, send the 
Bit Description
0 Unused
1 Request Bus Control: This bit is always set to 0. (The analyzer does not 
request control.)
2
Query Error: A 1 in this bit position indicates that a query error has 
occurred. Query errors have SCPI error numbers from 499 to 400.
3
Device Dependent Error: A 1 in this bit position indicates that a device 
dependent error has occurred. Device dependent errors have SCPI error numbers 
from –399 to –300 and 1 to 32767.
4 Execution Error: A 1 in this bit position indicates that an execution error 
has occurred. Execution errors have SCPI error numbers from –299 to –200.
5
Command Error: A 1 in this bit position indicates that a command error has 
occurred. Command errors have SCPI error numbers from –199 to –100.
6
User Request Key (Local): A 1 in this bit position indicates that the 
[Preset/System] (Local) key has been pressed. This is true even if the analyzer is 
in local lockout mode.
7
Power On: A 1 in this bit position indicates that the analyzer has been turned 
off and then on.