Technical data
Command Reference 5
Programmer’s Guide 85
Detector Subsection
Type  of  De te ction
[:SENSe]:DETector:TRACe[1]|2|3|4[:FUNCtion]NEGa-
tive|POSitive|SAMPle|AVERage|NORMAL
[:SENSe]:DETector:TRACe[1]|2|3|4[:FUNCtion]?
Specifies the detection mode. For each trace interval (bucket), average 
detection displays the average of all the samples within the interval.
• Negative peak detection displays the lowest sample taken during the 
interval being displayed.
• Positive peak detection displays the highest sample taken during the 
interval being displayed.
• Sample detection displays the sample taken during the interval being 
displayed, and is used primarily to display noise or noise-like signals. 
In sample mode, the instantaneous signal value at the present display 
point is placed into memory. This detection should not be used to 
make the most accurate amplitude measurement of non noise-like 
signals.
• Average detection is used when measuring the average value of the 
amplitude across each trace interval (bucket). The averaging method 
used by the average detector is set to either video or power as 
appropriate when the average type is auto coupled.
• Normal detection selects the maximum and minimum video signal 
values alternately. When selecting Normal detection, “Norm” 
appears in the upper-left corner.
*PST: Positive
Example: :DETector:TRACe1 SAMPle
Key access: Trace > More > Detector