Option H48 Multiport Test Set User's and Service Guide

Agilent Technologies Z5623A Option H48 5 -3
Performance Verification
Setting Test Limits
Applying Equation 5-1 to the uncertainties listed in Table 5-1 and to the
guaranteed specifications listed in Table 1-1 on page 1-4, we obtain the
test limits in Table 5-2:
Table 5-2 Derivation of Test Limits for Performance Verification of Test
Set
Parameter Specification Measurement
Uncertainty
Test Limit
Crosstalk
a,b
1) –110 dB
c
2) –105 dB
d
3) –95 dB
e
4) –90 dB
f
1) 5 dB
c
2) 5 dB
d
3) 5 dB
e
4) 7 dB
f
1) –115 dB
c
2) –110 dB
d
3) –100 dB
e
4) –97 dB
f
Return Loss
(port active)
g
1) 24 dB
c
2) 18 dB
d
3) 14 dB
e
4) 9 dB
f
1) 1.5 dB
c
2) 1.0 dB
d
3) 0.5 dB
e
4) 0.5 dB
f
1) 25.5 dB
c
2) 19 dB
d
3) 14.5 dB
e
4) 9.5 dB
f
Return Loss
(port off)
h
1) 26 dB
c
2) 20 dB
d
3) 16 dB
e
4) 14 dB
f
1) 1.5 dB
c
2) 1.5 dB
d
3) 1.0 dB
e
4) 0.5 dB
f
1) 27.5 dB
c
2) 21.5 dB
d
3) 17 dB
e
4) 14.5 dB
f
Insertion
Loss
i
1) 1.5 dB
c
2) 2.0 dB
d
3) 2.5 dB
e
4) 3.5 dB
f
0.3 dB
(all bands)
1) 1.8 dB
c
2) 2.3 dB
d
3) 2.8 dB
e
4) 3.8 dB
f
a. Between any two non-connected signal paths
b. Conditions: IF Bandwidth = 100 Hertz, Averaging = 10
c. Band 1 (300 kHz to 1.3 GHz)
d. Band 2 (1.3 GHz to 3.0 GHz)
e. Band 3 (3.0 GHz to 6.0 GHz)
f. Band 4 (6.0 GHz to 9.0 GHz)
g. When using an external termination on the Reflection or
Transmission port
h.When using one of the 50 internal terminations
i. From any test set port to the Reflection or Transmission port