Agilent 1100 Series Quaternary Pump Reference Manual sa
Copyright Agilent Technologies 1999 All rights reserved. Reproduction, adaption, or translation without prior written permission is prohibited, except as allowed under the copyright laws. Part No. G1311-90003 Edition 01/00 Printed in Germany Warranty WARNING The information contained in this document is subject to change without notice. For details of safety, see Safety Information on page 260.
Agilent 1100 Series Quaternary Pump Reference Manual
In This Book This manual contains technical reference information about the Agilent 1100 Series quaternary pump. The manual describes the following: • installation, • optimizing performance, • troubleshooting, • repairing, • parts and materials, • theory of operation, and • specifications.
Contents 1 Installing the Pump How to install the quaternary pump 13 Site Requirements 14 Unpacking the Quaternary Pump 17 Optimizing the Stack Configuration 20 Installing the Quaternary Pump 22 Flow Connections of the Quaternary Pump 25 Priming and Purging the System 28 2 Optimizing Performance How to optimize the quaternary pump to achieve best chromatographic results 31 Hints for Successful Use of the Quaternary Pump 32 Solvent Information 34 Prevent Blocking of Solvent Filters 35 Operational Hints for
Contents Power Supply Indicator 46 Pump Status Indicator 46 Error Messages 47 Timeout 48 Shut-Down 49 Remote Timeout 50 Synchronization Lost 51 Leak 52 Leak Sensor Open 53 Leak Sensor Short 54 Compensation Sensor Open 55 Compensation Sensor Short 56 Fan Failed 57 Open Cover 58 Restart Without Cover 59 Zero Solvent Counter 60 Pressure Above Upper Limit 61 Pressure Below Lower Limit 62 Pressure Signal Missing 63 Missing Pressure Reading 64 Pump Configuration 65 Valve Fuse 66 Inlet-Valve Fuse 67 Valve Failed
Contents Pump Head Missing 75 Index Limit 76 Index Adjustment 77 Index Missing 78 Stroke Length 79 Initialization Failed 80 Wait Timeout 81 Pressure Test 82 Running the Pressure Test 84 Evaluating the Results 86 Leak Test 88 Running the Leak Test 90 Evaluating the Results 92 4 Repairing the Pump Instructions on simple, routine repair procedures as well as more extensive repairs requiring exchange of internal parts 97 Cleaning the Quaternary Pump 99 Using the ESD Strap 100 Overview 101 Simple Repair Proced
Contents Removing and Disassembling the Pump Head Assembly 110 Exchanging the Pump Seals and Seal Wear-in Procedure 112 Exchanging the Plungers 115 Installing the Continuous Seal Wash Option 116 Exchanging the Wash Seals 119 Reassembling the Pump Head Assembly 121 Exchanging the Multi-Channel Gradient Valve (MCGV) 123 Exchanging the optional Interface Board 126 Exchanging Internal Parts 127 Removing the Top Cover and Foam 128 Exchanging the Low Pressure Pump Main Board (LPM Board) 131 Exchanging the Damper
Contents Bottle Head Assembly 165 Hydraulic Path 166 Cover Parts 167 Sheet Metal Kit 168 Foam Parts 169 Power and Status Light Pipes 170 Leak Parts 171 Pump Head Assembly 172 Pump Head Assembly with Seal Wash Option Outlet Ball Valve Assembly 176 Purge Valve Assembly 177 Active Inlet Valve Assembly 178 Accessory Kit G1311-68705 179 Seal Wash Option Kit 01018-68722 180 174 Cable Overview 181 Analog Cables 183 Remote Cables 185 BCD Cables 190 Auxiliary Cable 192 CAN Cable 192 External Contact Cable 193 RS-
Contents Introduction to the Quaternary Pump 198 Overview 199 Electrical Connections 204 Instrument Layout 206 Early Maintenance Feedback (EMF) 207 The Electronics 209 The Low-Pressure Pump Main Board (LPM) 210 Firmware Description 214 Optional Interface Boards 216 Agilent 1100 Series Interfaces 218 Setting the 8-bit Configuration Switch 223 The Main Power Supply Assembly 228 7 Control Module Screens for the Quaternary Pump Screens available from the Analysis screen 233 Screens available from the System s
Contents Radio Interference 264 Sound Emission 264 Solvent Information 265 Agilent Technologies on Internet 266 11
Contents 12
1 1 Installing the Pump How to install the quaternary pump
Installing the Pump Site Requirements Site Requirements A suitable environment is important to ensure optimum performance of the quaternary pump. Power Consideration The quaternary pump power supply has wideranging capability (see Table 1 on page 16). It accepts any line voltage in the range described in the above mentioned table. Consequently there is no voltage selector in the rear of the quaternary pump.
Installing the Pump Site Requirements Bench Space The quaternary pump dimensions and weight (see Table 1 on page 16) allow to place the quaternary pump on almost any laboratory bench. It needs an additional 2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches) in the rear for the circulation of air and electric connections. If the bench should carry a complete Agilent 1100 Series system, make sure that the bench is designed to carry the weight of all the modules.
Installing the Pump Site Requirements Table 1 Physical Specifications Type Specification Weight 11 kg (25 lbs) Dimensions (height × weight × depth) 140 × 345 × 435 mm (5.5 × 13.
Installing the Pump Unpacking the Quaternary Pump Unpacking the Quaternary Pump Damaged Packaging Upon receipt of your quaternary pump, inspect the shipping containers for any signs of damage. If the containers or cushioning material are damaged, save them until the contents have been checked for completeness and the quaternary pump has been mechanically and electrically checked.
Installing the Pump Unpacking the Quaternary Pump Table 2 Quaternary Pump Checklist, continued Description Quantity CAN cable 1 Remote cable As ordered Signal cable As ordered Reference Manual 2 (1 for the pump, 1 for the degasser) Accessory kit (see Table 3) 1 Accessory Kit Contents — Quaternary Pump Table 3 Accessory Kit Contents G1311-68705 Description Part Number Quantity Capillary, pump to injection device G1312-67305 1 Seal insert tool 01018-23702 1 Wrench; 1/4 – 5/16 inch 8710
Installing the Pump Unpacking the Quaternary Pump Accessory Kit Contents—Vacuum Degasser Table 4 Accessory Kit Contents G1322-68705 Description Part Number Quantity Syringe 5062-8534 1 Syringe adapter 9301-1337 1 Waste tube (reorder number, 5m) 5062-2463 1.
Installing the Pump Optimizing the Stack Configuration Optimizing the Stack Configuration If your quaternary pump is part of a complete 1100 Series system, you can ensure optimum performance by limiting the configuration of the system stack to the following configuration. This configuration optimizes the system flow path, ensuring minimum delay volume. Figure 1 Recommended Stack Configuration (Front View) Flow connections in the stack: Example setup with 0.
Installing the Pump Optimizing the Stack Configuration Figure 2 Recommended Stack Configuration (Rear View) Remote cable 5061-3378 Pressure output to recorder, for PN see page 181 CAN Bus cable to handheld controller G1323-81600 CAN Bus cable for inter module communication 5181-1516 (0.5m) 5161-1519 (1.0m) AC power Analog signal to recorder,for PN see page 181 NOTE GPIB or LAN to ChemStation for PN see page 181 If a single stack configuration becomes too high, e.g.
Installing the Pump Installing the Quaternary Pump Installing the Quaternary Pump Preparations Locate bench space. Provide power connections. Unpack the pump. Parts required Pump Power cord, for other cables see text below and “Cable Overview” on page 181 ChemStation and/or Control Module G1323A/B 1 Place the quaternary pump on the bench in a horizontal position. 2 Ensure the power switch on the front of the quaternary pump is OFF (switch stands out).
Installing the Pump Installing the Quaternary Pump NOTE In an Agilent 1100 Series system, the individual modules are connected through CAN cables. The Agilent 1100 Series vacuum degasser is an exception . The vacuum degasser can be connected via the APG remote connector to the other modules of the stack. The AUX output allows the user to monitor the vacuum level in the degasser chamber.
Installing the Pump Installing the Quaternary Pump NOTE The power switch stays pressed in and a green indicator lamp in the power switch is on when the quaternary pump is turned on. When the line power switch stands out and the green light is off, the quaternary pump is turned off. 8 Purge the quarternary pump (see “Priming and Purging the System” on page 28). WAR NI N G To disconnect the quaternary pump from line, unplug the power cord.
Installing the Pump Flow Connections of the Quaternary Pump Flow Connections of the Quaternary Pump Preparations Pump is installed in the LC system. Parts required Other modules Parts from accessory kit, see “Accessory Kit Contents — Quaternary Pump” on page 18 Two wrenches 1/4–5/16 inch for capillary connections WAR NI N G When opening capillary or tube fittings solvents may leak out.
Installing the Pump Flow Connections of the Quaternary Pump 5 Connect the solvent tubes to the outlet connectors (typically right connection of the channel) of the vacuum degasser. 6 Connect the syringe adapter from the degasser accessory kit to the solvent tube of channel A. 7 Using a piece of sanding paper connect the waste tubing to the purge valve and place it into your waste system.
Installing the Pump Flow Connections of the Quaternary Pump Figure 6 Flow Connections of the Quaternary Pump Bottle-head assembly (G1311-60003) Solvent cabinet Vacuum degasser Inlet Outlet Tube clip (1400-1578) Tubings (G1322-67300) Purge valve Waste tubing (5062-2461) Outlet capillary to autosampler (G1312-67305) MCGV Fitting for corrugated waste tubing (5062-2463, reorder pack, 5m) 27
Installing the Pump Priming and Purging the System Priming and Purging the System The system can be primed either by drawing solvent through the degasser with a syringe or by pumping with the pump. Priming the system with a syringe is recommended, when: • vacuum degasser or connected tubings are used for the first time or vacuum tubes are empty or • changing to solvents that are immiscible with the solvent currently in the vacuum tubes.
Installing the Pump Priming and Purging the System 5 Push syringe adapter onto syringe. 6 Pull syringe plunger to draw at least 30 ml of solvent through degasser and tubing. 7 Replace the priming solvent with the new solvent of your choice. 8 Pull syringe plunger to draw at least 30 ml of solvent through degasser and tubing. 9 Disconnect syringe adapter from solvent tube. 10 Connect the solvent tube to the appropriate channel of the MCGV. 11 Repeat step 3 to step 10 for the other solvent channels.
Installing the Pump Priming and Purging the System Table 5 Choice of Priming Solvents for Different Purposes Activity Solvent Comments After an installation Isopropanol Best solvent to flush air out of the system When switching between reverse phase and normal phase (both times) Isopropanol Best solvent to flush air out of the system After an installation Ethanol or Methanol Alternative to Isopropanol (second choice) if no Isopropanol is available To clean the system when using buffers Bidist
2 2 Optimizing Performance How to optimize the quaternary pump to achieve best chromatographic results
Optimizing Performance Hints for Successful Use of the Quaternary Pump Hints for Successful Use of the Quaternary Pump • Always place the solvent cabinet with the solvent bottles on top of the quaternary pump (or at a higher level). • When using salt solutions and organic solvents in the Agilent 1100 Quaternary Pump it is recommended to connect the salt solution to one of the bottom gradient valve ports and the organic solvent to one of the upper gradient valve ports.
Optimizing Performance Hints for Successful Use of the Quaternary Pump Procedure” on page 112).
Optimizing Performance Solvent Information Solvent Information Always filter solvents through 0.4 µm filters, small particles can permanently block the capillaries and valves. Avoid the use of the following steel-corrosive solvents: • Solutions of alkali halides and their respective acids (for example, lithium iodide, potassium chloride, and so on).
Optimizing Performance Prevent Blocking of Solvent Filters Prevent Blocking of Solvent Filters Contaminated solvents or algae growth in the solvent bottle will reduce the lifetime of the solvent filter and will influence the performance of the pump. This is especially true for aqueous solvents or phosphate buffers (pH 4 to 7).
Optimizing Performance Operational Hints for the Vacuum Degasser Operational Hints for the Vacuum Degasser Operational Hints for the Vacuum Degasser If you are using the vacuum degasser for the first time, if the vacuum degasser was switched off for any length of time (for example, overnight), or if the vacuum degasser lines are empty, you should prime the vacuum degasser before running an analysis.
Optimizing Performance Operational Hints for the Multi Channel Gradient Valve (MCGV) Operational Hints for the Multi Channel Gradient Valve (MCGV) In a mixture of salt solutions and organic solvent the salt solution might be well dissolved in the organic solvent without showing precipitations. However in the mixing point of the gradient valve, at the boundary between the two solvents, micro precipitation is possible. Gravity forces the salt particles to fall down.
Optimizing Performance When to use the Continuous Seal Wash Option When to use the Continuous Seal Wash Option Highly-concentrated buffer solutions will reduce the lifetime of the seals and plungers in your quaternary pump. The seal wash option allows to maintain the seal lifetime by flushing the back side of the seal with a wash solvent. The continuous seal wash option is strongly recommended when buffer concentrations of 0.1 Molar or higher will be used for long time periods in the quaternary pump.
Optimizing Performance When to Use Alternative Seals When to Use Alternative Seals The standard seal for the quaternary pump can be used for most applications. However applications that use normal phase solvents (for example, hexane) are not suited for the standard seal and require a different seal when used for a longer time in the quaternary pump. For applications that use normal phase solvents (for example, hexane) we recommend the use of the polyethylene seals, part number 0905-1420 (pack of 2).
Optimizing Performance Optimize the Compressibility Compensation Setting Optimize the Compressibility Compensation Setting The compressibility compensation default setting is 100 × 10-6 /bar for the quaternary pump. This setting represents an average value. Under normal conditions the default setting reduces the pressure pulsation to values (below 1% of system pressure) that will be sufficient for most applications and for all gradient analyses.
Optimizing Performance Optimize the Compressibility Compensation Setting 5 Starting with a compressibility setting of 10 × 10-6 /bar increase the value in steps of 10. Re-zero the integrator as required. The compressibility compensation setting that generates the smallest pressure ripple is the optimum value for your solvent composition.
Optimizing Performance Optimize the Compressibility Compensation Setting 42
3 3 Troubleshooting and Test Functions The quaternary pump’s built-in troubleshooting and test functions
Troubleshooting and Test Functions This chapter describes the instrument’s built in troubleshooting and test functions. Status Indicators The quaternary pump is provided with two status indicators which indicate the operational state (prerun, run, and error states) of the quaternary pump. The status indicators provide a quick visual check of the operation of the quaternary pump (see “Status Indicators” on page 45).
Troubleshooting and Test Functions Status Indicators Two status indicators are located on the front of the quaternary pump. The lower left one indicates the power supply status, the upper right one indicates the quaternary pump status.
Power Supply Indicator The power supply indicator is integrated into the main power switch. When the indicator is illuminated (green) the power is ON. When the indicator is off, the module is turned OFF. Otherwhise check power connections, availability of power or check functioning of the power supply.
Troubleshooting and Test Functions Error Messages Error Messages Error messages are displayed in the user interface when an electronic, mechanical, or hydraulic (flow path) failure occurs which requires attention before the analysis can be continued (for example, repair, frit exchange or exchange of consumables required). In the event of such a failure, the red status indicator at the front of the quaternary pump is switched on, and an entry is written into the instrument logbook.
Timeout The timeout threshold was exceeded. Probable Causes • The analysis was completed successfully, and the timeout function switched off the quaternary pump as requested. • A not-ready condition was present during a sequence or multiple-injection run for a period longer than the timeout threshold. Suggested Actions ❏ Check the logbook for the occurrence and source of a not-ready condition. Restart the analysis where required.
Troubleshooting and Test Functions Shut-Down Shut-Down An external instrument has generated a shut-down signal on the remote line. The quaternary pump continually monitors the remote input connectors for status signals. A LOW signal input on pin 4 of the remote connector generates the error message. Probable Causes • Leak detected in another module with a CAN connection to the system . • Leak detected in an external instrument with a remote connection to the system.
Troubleshooting and Test Functions Remote Timeout Remote Timeout A not-ready condition is still present on the remote input . When an analysis is started, the system expects all not-ready conditions (e.g. a not-ready condition during detector balance) to switch to run conditions within one minute of starting the analysis. If a not-ready condition is still present on the remote line after one minute the error message is generated.
Troubleshooting and Test Functions Synchronization Lost Synchronization Lost During an analysis, the internal synchronization or communication between one or more of the modules in the system has failed. The system processors continually monitor the system configuration. If one or more of the modules is no longer recognized as being connected to the system, the error message is generated. Probable Causes • CAN cable disconnected. • Defective CAN cable. • Defective main board in another module.
Troubleshooting and Test Functions Leak Leak A leak was detected in the quaternary pump. The signals from the two temperature sensors (leak sensor and board-mounted temperature-compensation sensor) are used by the leak algorithm to determine whether a leak is present. When a leak occurs, the leak sensor is cooled by the solvent. This changes the resistance of the leak sensor which is sensed by the leak-sensor circuit on the LPM board Probable Causes • Loose fittings. • Broken capillary.
Troubleshooting and Test Functions Leak Sensor Open Leak Sensor Open The leak sensor in the quaternary pump has failed (open circuit). The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak-sensor current to change within defined limits. If the current falls outside the lower limit, the error message is generated. Probable Causes • Leak sensor not connected to the LPM board. • Defective leak sensor.
Troubleshooting and Test Functions Leak Sensor Short Leak Sensor Short The leak sensor in the quaternary pump has failed (short circuit). The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak-sensor current to change within defined limits. If the current increases above the upper limit, the error message is generated. Probable Causes • Defective leak sensor.
Troubleshooting and Test Functions Compensation Sensor Open Compensation Sensor Open The ambient-compensation sensor (NTC) on the LPM board in the quaternary pump has failed (open circuit). The resistance across the temperature compensation sensor (NTC) on the LPM board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor increases above the upper limit, the error message is generated.
Troubleshooting and Test Functions Compensation Sensor Short Compensation Sensor Short The ambient-compensation sensor (NTC) on the LPM board in the quaternary pump has failed (short circuit). The resistance across the temperature compensation sensor (NTC) on the LPM board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor falls below the lower limit, the error message is generated.
Troubleshooting and Test Functions Fan Failed Fan Failed The cooling fan in the quaternary pump has failed. The hall sensor on the fan shaft is used by the LPM board to monitor the fan speed. If the fan speed falls below 2 revolutions/second for longer than 5 seconds, the error message is generated. Probable Causes • Fan cable disconnected. • Defective fan. • Defective LPM board. • Improperly positioned cables or wires obstructing fan blades. Suggested Actions ❏ Ensure the fan is connected correctly.
Troubleshooting and Test Functions Open Cover Open Cover The top foam has been removed. The sensor on the LPM board detects when the top foam is in place. If the foam is removed, the fan is switched off, and the error message is generated. Probable Causes • The top foam was removed during operation. • Foam not activating the sensor. • Sensor defective. • Rear of the module is exposed to strong direct sunlight. Suggested Actions ❏ Replace the top foam. ❏ Exchange the LPM board.
Troubleshooting and Test Functions Restart Without Cover Restart Without Cover The quaternary pump was restarted with the top cover and foam open. The sensor on the LPM board detects when the top foam is in place. If the quaternary pump is restarted with the foam removed, the quaternary pump switches off within 30 s, and the error message is generated. Probable Causes • The quaternary pump started with the top cover and foam removed. • Rear of the module is exposed to strong direct sunlight.
Troubleshooting and Test Functions Zero Solvent Counter Zero Solvent Counter Pump firmware version A.02.32 and higher allow to set solvent bottle fillings at the ChemStation (revision 5.xx and higher). If the volume level in the bottle falls below the specified value the error message appears when the feature is configured accordingly. Probable Causes • Volume in bottle below specified volume. • Incorrect setting of limit. Suggested Actions ❏ Refill bottles and reset solvent counters.
Troubleshooting and Test Functions Pressure Above Upper Limit Pressure Above Upper Limit The system pressure has exceeded the upper pressure limit. Probable Causes • Upper pressure limit set too low. • Blockage in the flowpath (after the damper). • Defective damper. • Defective LPM board. Suggested Actions ❏ Ensure the upper pressure limit is set to a value suitable for the analysis. ❏ Check for blockage in the flowpath. The following components are particularly subject to blockage: purge-valve frit.
Troubleshooting and Test Functions Pressure Below Lower Limit Pressure Below Lower Limit The system pressure has fallen below the lower pressure limit. Probable Causes • Lower pressure limit set too high. • Air bubbles in the mobile phase. • Leak. • Defective damper. • Defective LPM board. Suggested Actions ❏ Ensure the lower pressure limit is set to a value suitable for the analysis. ❏ Ensure solvents are degassed. Purge the quaternary pump. ❏ Ensure solvent inlet filters are not blocked.
Troubleshooting and Test Functions Pressure Signal Missing Pressure Signal Missing The pressure signal from the damper is missing. The pressure signal from the damper must be within a specific voltage range. If the pressure signal is missing, the processor detects a voltage of approximately -120mV across the damper connector. Probable Causes • Damper disconnected. • Defective damper. Suggested Actions ❏ Ensure the damper is connected correctly to the LPM board. ❏ Exchange the damper.
Troubleshooting and Test Functions Missing Pressure Reading Missing Pressure Reading The pressure readings read by the pump ADC (analog-digital converter) are missing. The ADC reads the pressure readings from the damper every 1ms. If the readings are missing for longer than 10 seconds, the error message is generated. Probable Causes • Damper not connected. • Defective damper. • Defective LPM board. Suggested Actions ❏ Ensure the damper is connected, clean and seated correctly. ❏ Exchange the damper.
Troubleshooting and Test Functions Pump Configuration Pump Configuration At switch-on, the quaternary pump has recognized a new pump configuration. The quaternary pump is assigned its configuration at the factory. If the gradient valve is disconnected, and the quaternary pump is rebooted, the error message is generated. However, the pump will function as an isocratic pump in this configuration.The error message reappears after each switch-on. Probable Causes • Gradient valve disconnected.
Troubleshooting and Test Functions Valve Fuse Valve Fuse Valve Fuse 0: Channels A and B Valve Fuse 1: Channels C and D The gradient valve in the quaternary pump has drawn excessive current causing the electronic fuse to open. Probable Causes • Defective gradient valve. • Defective connection cable (front panel to LPM board). • Defective LPM board. Suggested Actions ❏ Restart the quaternary pump. If the error message appears again, exchange the gradient valve. ❏ Exchange the connection cable.
Troubleshooting and Test Functions Inlet-Valve Fuse Inlet-Valve Fuse The active-inlet valve in the quaternary pump has drawn excessive current causing the inlet-valve electronic fuse to open. Probable Causes • Defective active inlet valve. • Defective connection cable (front panel to LPM board). • Defective LPM board. Suggested Actions ❏ Restart the quaternary pump. If the error message appears again, exchange the active inlet valve. ❏ Exchange the connection cable. ❏ Exchange the LPM board.
Troubleshooting and Test Functions Valve Failed Valve Failed Valve 0 Failed: valve A Valve 1 Failed: valve B Valve 2 Failed: valve C Valve 3 Failed: valve D One of the valves of the multi-channel gradient valve has failed to switch correctly. The processor monitors the valve voltage before and after each switching cycle. If the voltages are outside expected limits, the error message is generated. Probable Causes • Gradient valve disconnected. • Connection cable (inside instrument) not connected.
Troubleshooting and Test Functions Motor-Drive Power Motor-Drive Power The current drawn by the pump motor exceeded the maximum limit. Blockages in the flow path are usually detected by the pressure sensor in the damper, which result in the pump switching off when the upper pressure limit is exceeded. If a blockage occurs before the damper, the pressure increase cannot be detected by the pressure sensor and the quaternary pump will continue to pump. As pressure increases, the pump drive draws more current.
Troubleshooting and Test Functions Encoder Missing Encoder Missing The optical encoder on the pump motor in the quaternary pump is missing or defective. The processor checks the presence of the pump encoder connector every 2 seconds. If the connector is not detected by the processor, the error message is generated. Probable Causes • Defective or disconnected pump encoder connector. • Defective pump drive assembly. Suggested Actions ❏ Ensure the connector is clean, and seated correctly.
Troubleshooting and Test Functions Inlet-Valve Missing Inlet-Valve Missing The active-inlet valve in the quaternary pump is missing or defective. The processor checks the presence of the active-inlet valve connector every 2 seconds. If the connector is not detected by the processor, the error message is generated. Probable Causes • Disconnected or defective cable. • Disconnected or defective connection cable (front panel to LPM board). • Defective active inlet valve.
Troubleshooting and Test Functions Temperature Out of Range Temperature Out of Range The temperature sensor readings in the motor-drive circuit are out of range. The values supplied to the ADC by the hybrid sensors must be between 0.5 V and 4.3 V. If the values are outside this range, the error message is generated. Probable Causes • Defective LPM board. Suggested Actions ❏ Exchange the LPM board.
Troubleshooting and Test Functions Temperature Limit Exceeded Temperature Limit Exceeded The temperature of one of the motor-drive circuits is too high. The processor continually monitors the temperature of the drive circuits on the LPM board. If excessive current is being drawn for long periods, the temperature of the circuits increases. If the temperature exceeds the upper limit of 95 ºC, the error message is generated.
Troubleshooting and Test Functions Servo Restart Failed Servo Restart Failed The pump motor in the quaternary pump was unable to move into the correct position for restarting. When the quaternary pump is switched on, the first step is to switch on the C phase of the variable reluctance motor. The rotor should move to one of the C positions. The C position is required for the servo to be able to take control of the phase sequencing with the commutator.
Troubleshooting and Test Functions Pump Head Missing Pump Head Missing The pump-head end stop in the quaternary pump was not found. When the quaternary pump restarts, the metering drive moves forward to the mechanical end stop. Normally, the end stop is reached within 20 seconds, indicated by an increase in motor current. If the end point is not found within 20 seconds, the error message is generated.
Troubleshooting and Test Functions Index Limit Index Limit The time required by the plunger to reach the encoder index position was too short (quaternary pump). During initialization, the first plunger is moved to the mechanical stop. After reaching the mechanical stop, the plunger reverses direction until the encoder index position is reached. If the index position is reached too fast, the error message is generated. Probable Causes • Irregular or sticking drive movement. • Defective pump drive assembly.
Troubleshooting and Test Functions Index Adjustment Index Adjustment The encoder index position in the quaternary pump is out of adjustment. During initialization, the first plunger is moved to the mechanical stop. After reaching the mechanical stop, the plunger reverses direction until the encoder index position is reached. If the time to reach the index position is too long, the error message is generated. Probable Causes • Irregular or sticking drive movement. • Defective pump drive assembly.
Troubleshooting and Test Functions Index Missing Index Missing The encoder index position in the quaternary pump was not found during initialization. During initialization, the first plunger is moved to the mechanical stop. After reaching the mechanical stop, the plunger reverses direction until the encoder index position is reached. If the index position is not recognized within a defined time, the error message is generated. Probable Causes • Disconnected or defective encoder cable.
Troubleshooting and Test Functions Stroke Length Stroke Length The distance between the lower plunger position and the upper mechanical stop is out of limits (quaternary pump). During initialization, the quaternary pump monitors the drive current. If the plunger reaches the upper mechanical stop position before expected, the motor current increases as the quaternary pump attempts to drive the plunger beyond the mechanical stop. This current increase causes the error message to be generated.
Troubleshooting and Test Functions Initialization Failed Initialization Failed The quaternary pump failed to initialize successfully within the maximum time window. A maximum time is assigned for the complete pump-initialization cycle. If the time is exceeded before initialization is complete, the error message is generated. Probable Causes • Blocked active inlet valve. • Defective pump drive assembly. • Defective LPM board. Suggested Actions ❏ Exchange the active inlet valve.
Troubleshooting and Test Functions Wait Timeout Wait Timeout When running certain tests in the diagnostics mode or other special applications, the pump must wait for the plungers to reach a specific position, or must wait for a certain pressure or flow to be reached. Each action or state must be completed within the timeout period, otherwise the error message is generated. Possible Reasons for a Wait Timeout • Pressure not reached. • Pump channel A did not reach the delivery phase.
Pressure Test Description The pressure test is a quick, built-in test designed to demonstrate the pressure-tightness of the system. The test should be used when problems with small leaks are suspected, or after maintenance of flow-path components (e.g. pump seals, injection seal) to prove pressure tightness up to 400 bar. The test involves monitoring the pressure profile as the pump runs through a predefined pumping sequence.
Troubleshooting and Test Functions Pressure Test with a flow rate of 510 µl/min and stroke of 100 µl. The pump continues to pump until a system pressure of 390 bar is reached. Step 2 When the system pressure reaches 390 bar, the pump switches off. The pressure drop from this point onwards should be no more than 2 bar/minute.
Troubleshooting and Test Functions Running the Pressure Test Running the Pressure Test Tools required ¼-inch” wrench Parts and materials required Blank nut, 01080-83202 Isopropanol, 500 ml Running the test from the ChemStation 1 Select the pressure test from the test selection box in the Diagnosis screen. 2 Start the test and follow the instructions.
Troubleshooting and Test Functions Running the Pressure Test Once the test is started, the pump moves the plungers into the start position. When the plungers are in position, the user interface prompts you to close the purge valve, and continue the test. 7 Close the purge valve, select continue on the control module and press Enter to start the test. The control module displays a graphical representation of the pressure.
Troubleshooting and Test Functions Evaluating the Results Evaluating the Results The sum of all leaks between the pump and the blank nut will be indicated by a pressure drop of >2 bar/minute at the plateau. Note that small leaks may cause the test to fail, but solvent may not be seen leaking from a module. NOTE Please notice the difference between an error in the test and a failure of the test! An error means that during the operation of the test there was an abnormal termination.
Troubleshooting and Test Functions Evaluating the Results Potential Causes of Pressure Test Failure After isolating and fixing the cause of the leak, repeat the pressure test to confirm the system is pressure tight. Potential Cause (Pump) Corrective Action Purge valve open. Close the purge valve. Loose or leaky fitting. Tighten the fitting or exchange the capillary. Damaged pump seals or plungers. Run the leak test to confirm the leak. Loose purge valve. Tighten the purge valve nut (14 mm wrench).
Troubleshooting and Test Functions Leak Test Leak Test Description The leak test is a built-in troubleshooting test designed to demonstrate the leak-tightness of the pump. The test should be used when problems with the pump are suspected. The test involves monitoring the pressure profile as the pump runs through a predefined pumping sequence. The resulting pressure profile provides information about the pressure tightness and operation of the pump components.
Troubleshooting and Test Functions Leak Test Ramp 1: After initialization, plunger 2 is at the top of its stroke. The test begins with plungerplunger 1 delivering with a stroke length of 100µl and a flow of 153µl/min. The plunger sequence during the pressure ramp is 1-2-1-2. The pressure increase during this phase should be linear. Pressure disturbances during this phase indicate larger leaks or defective pump components.
Running the Leak Test Tools required ¼ inch” wrench. Parts and materials required Restriction Capillary ,G1313-87305 Blank nut, 01080-83202 Isopropanol, 500 ml Running the test from the ChemStation 1 Select the leak test from the test selection box in the Diagnosis screen. 2 Start the test and follow the instructions.
Troubleshooting and Test Functions Running the Leak Test 9 Connect the signal cable to the analog output at the rear of the pump module (only if an integrator is used). 10 Press Execute to initialize the leak test. Once the test is started, the pump moves the plungers into the start position. When the plungers are in position, the user interface prompts you to close the purge valve. 11 Close the purge valve, select continue on the control module and press Enter to start the test.
Troubleshooting and Test Functions Evaluating the Results Evaluating the Results Defective or leaky components in the pump head lead to changes in the leak-test pressure plot. Typical failure modes are described below.
Troubleshooting and Test Functions Evaluating the Results No Pressure increase at Ramp 1 Potential Cause Corrective Action Pump not running. Check the logbook for error messages. Purge valve open. Close the purge valve, and restart the test. Loose or leaky fittings. Ensure all fittings are tight, or exchange capillary. Wrong solvent-line connections. Ensure the solvent lines from the degasser are connected correctly (channel D for quaternary pump). Contaminated purge valve.
Troubleshooting and Test Functions Evaluating the Results All plateaus negative Potential Cause Corrective Action Loose or leaky fittings. Ensure all fittings are tight, or exchange capillary. Loose purge valve. Tighten the purge valve (14mm wrench). Contaminated purge valve. Open and close purge valve to flush out contamination. Exchange the valve if still leaky. Loose pump head screws. Ensure the pump head screws are tight. Leaking seals or scratched plungers. Exchange the pump seals.
Troubleshooting and Test Functions Evaluating the Results First plateau negative, second plateau positive Potential Cause Corrective Action Leaking outlet valve. Clean the outlet valve. Ensure the sieve in the outlet valves are installed correctly. Tighten the outlet valve. Loose pump head screws. Ensure the pump head screws are tight. Leaking seals or scratched plungers. Exchange the pump seals. Check the plunger for scratches. Exchange if scratched.
Troubleshooting and Test Functions Evaluating the Results 96
4 4 Repairing the Pump Instructions on simple, routine repair procedures as well as more extensive repairs requiring exchange of internal parts
Repairing the Pump Simple Repairs The quaternary pump is designed for easy repair. The most frequent repairs such as plunger seal change and purge valve frit change can be done from the front of the quaternary pump with the quaternary pump in place in the system stack. These repairs are described in “Simple Repair Procedures” on page 102. WAR NI N G When opening capillary or tube fittings solvents may leak out.
Repairing the Pump Cleaning the Quaternary Pump Cleaning the Quaternary Pump The quaternary pump case should be kept clean. Cleaning should be done with a soft cloth slightly dampened with water or a solution of water and a mild detergent. Do not use an excessively damp cloth that liquid can drip into the quaternary pump. WAR NI N G Do not let liquid drip into the quaternary pump. It could cause shock hazard and it could damage the quaternary pump.
Repairing the Pump Using the ESD Strap Using the ESD Strap Electronic boards are sensitive to electrostatic discharge (ESD). In order to prevent damage, always use an ESD strap supplied in the standard accessory kit (see “Accessory Kit G1311-68705” on page 179) when handling electronic boards and components. Using the ESD Strap 1 Unwrap the first two folds of the band and wrap the exposed adhesive side firmly around your wrist.
Repairing the Pump Overview Overview Figure 12 shows the main assemblies of the quaternary pump. The pump head and its parts do require normal maintenance (for example, seal exchange) and can be accessed from the front (simple repairs). Replacing internal parts will require to remove the quaternary pump from its stack and to open the top cover.
Simple Repair Procedures The procedures described in this section can be done with the quaternary pump in place in the system stack.
Repairing the Pump Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve When required If internally leaking (backflow) Tools required Wrench 14 mm Pair of Tweezers Material Active inlet valve G1312-60010 (complete assembly) Valve cartridge 5062-8562 Removing the Active Inlet Valve 1 Unplug the active inlet valve cable from the connector. 2 Disconnect the solvent inlet tube at the inlet valve.
Repairing the Pump Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve Figure 13 Active Inlet Valve Assembly Valve body Valve cartridge Replacing the Active Inlet Valve 1 Insert the new, or rebuilt valve into the pump head. Using the 14 mm wrench turn the nut until hand tight. 2 Position the valve that the solvent inlet tube connection points towards the front. 3 Using the 14 mm wrench tighten the nut by turning the valve in its final position (not more than a quarter turn).
Repairing the Pump Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve Figure 14 Exchanging the Active Inlet Valve Capillary pump to injection device Connector Solvent inlet tube Active inlet valve cable Active inlet valve 105
Repairing the Pump Exchanging the Outlet Ball Valve Exchanging the Outlet Ball Valve When required If internally leaking Tools required Wrench 1/4 inch Wrench 14 mm Parts required Outlet ball valve G1311-60012 NOTE Before exchanging the outlet ball valve you can try to clean it in a sonic bath. Place the valve in upright position (onto the plastic cap) in a small beaker with alcohol. Place in a sonic bath for 5 – 10 minutes.
Repairing the Pump Exchanging the Outlet Ball Valve Figure 16 Exchanging the Outlet Ball Valve Valve capillary Outlet ball valve 107
Repairing the Pump Exchanging the Purge Valve Frit or the Purge Valve Exchanging the Purge Valve Frit or the Purge Valve When required Frit – when plunger seals are exchanged or when contaminated or blocked (pressure drop of > 10 bar across the frit at a flow rate of 5 ml/min of H20 with purge valve opened) Purge valve – if internally leaking Tools required Wrench 1/4 inch Wrench 14 mm Pair of tweezers or toothpick Parts required PFTE frit (pack of 5) 01018-22707 Purge valve G1311-60009 1 Using a 1/4
Repairing the Pump Exchanging the Purge Valve Frit or the Purge Valve 6 Place a new frit into the purge valve with the orientation of the frit as shown above. 7 Reinstall the cap and the gold seal. NOTE Before reinstallation always check the gold seal. A deformed seal should be exchanged. 8 Insert the purge valve into the pump head and locate the pump outlet capillary and the waste tube as shown in Figure 18. 9 Tighten the purge valve and reconnect outlet capillary and waste tubing.
Repairing the Pump Removing and Disassembling the Pump Head Assembly Removing and Disassembling the Pump Head Assembly WAR NI N G Never start the pump when the pump head is removed. This may damage the pump drive.
Repairing the Pump Removing and Disassembling the Pump Head Assembly 3 Place the pump head on a flat surface. Loosen the lock screw (two revolutions) and while holding the lower half of the assembly carefully pull the pump head away from the plunger housing. Pump head Lock screw Plunger housing 4 Remove the support rings from the plunger housing and lift the housing away from the plungers.
Repairing the Pump Exchanging the Pump Seals and Seal Wear-in Procedure Exchanging the Pump Seals and Seal Wear-in Procedure When required: ❏Seal leaking, if indicated by the results of the leak test 1 Disassemble the pump head assembly (see “Removing and Disassembling the Pump Head Assembly” on page 110).
Repairing the Pump Exchanging the Pump Seals and Seal Wear-in Procedure 4 Insert new seals into the pump head. 5 Reassemble the pump head assembly (see “Reassembling the Pump Head Assembly” on page 121). Reset seal wear counter and liquimeter as described in the documentation. Seals Seal Wear-in Procedure NOTE This procedure is required for standard seals only (5063-6589), but it will definetely damage the normal phase application seals (0905-1420).
Repairing the Pump Exchanging the Pump Seals and Seal Wear-in Procedure 6 Turn OFF the pump, slowly open the purge valve to release the pressure from the system, disconnect the restriction capillary and reconnect the outlet capillary at the purge valve and the connecting tube from MCGV to the AIV. 7 Rinse your system with the solvent used for your next application.
Repairing the Pump Exchanging the Plungers Exchanging the Plungers 1 Disassemble the pump head assembly (see “Removing and Disassembling the Pump Head Assembly” on page 110). When required: ❏When scratched Tools required: ❏4-mm hexagonal key Parts: ❏Plunger 5063-6586 2 Check the plunger surface and remove any deposits or layers. Cleaning can be done with alcohol or tooth paste. Replace plunger if scratched.
Repairing the Pump Installing the Continuous Seal Wash Option Installing the Continuous Seal Wash Option 1 Disconnect all capillaries and tubes from the pump head and disconnect the active inlet valve cable. Tools required: ❏4-mm hexagonal key Parts: ❏Seal wash kit (01018-68722) 2 Using a 4-mm hexagonal key stepwise loosen and remove the two pump head screws and remove the pump head from the pump drive. 3 Place the pump head on a flat surface.
Repairing the Pump Installing the Continuous Seal Wash Option 4 Remove the support rings from the plunger housing and lift the housing away from the plungers. 5 Check the plunger surface and remove any deposits or layers. Cleaning can be done with alcohol or tooth paste. Replace plunger if scratched. Support ring Plunger surface Plunger housing Plunger 6 Install the support ring assembly from the seal wash option kit into the plunger housing. 7 If necessary replace the pump seals (see page 112 ).
Repairing the Pump Installing the Continuous Seal Wash Option 1 Route the wash inlet tube into a bottle filled with a mixture of distilled water and isopropanol (90/10) and place the bottle above the pump (hydrostatic pressure) in the solvent cabinet. 2 Route the outlet of the wash tube into a waste container. 3 The flow rate should be set to approximately 20 drops/minute. Use the velocity regulator attached to the wash tube to regulate the flow rate. NOTE The seals should never run dry.
Repairing the Pump Exchanging the Wash Seals Exchanging the Wash Seals 1 Disassemble the pump head assembly (see “Removing and Disassembling the Pump Head Assembly” on page 110). Tools required: ❏4-mm hexagonal key ❏Insert tool Parts: ❏Wash Seal 0905-1175 ❏Gasket, seal wash (pack of 6) 5062-2484 2 Remove the seal keeper and the seal wash support rings from the plunger housing. Remove the seal keeper from the support ring assembly.
Repairing the Pump Exchanging the Wash Seals 4 Using the insert tool press the secondary seal (spring pointing upwards) into the recess of the support ring. Place a seal wash gasket in the recess of the support ring. Insert tool Seal Support ring 120 5 Reassemble the pump head assembly (see “Reassembling the Pump Head Assembly” on page 121).
Repairing the Pump Reassembling the Pump Head Assembly Reassembling the Pump Head Assembly 1 Place the support rings on the plunger housing (plungers not installed) and snap the pump head and plunger housing together. Tools required: ❏4-mm hexagonal key ❏Pump head grease (79846-65501) Pump head Support ring Plunger housing 2 Insert the plungers and carefully press them into the seal. 3 Tighten the lock screw.
Repairing the Pump Reassembling the Pump Head Assembly 4 Slide the pump head assembly onto the metering drive. Apply a small amount of pump head grease to the pumphead screws and the balls of the spindle drive. Tighten the pumphead screws stepwise with increasing torque. Metering drive Balls of spindle drive Pump head Pumphead screws 122 5 Reconnect all capillaries, tubes and the active inlet valve cable to its connector.
Repairing the Pump Exchanging the Multi-Channel Gradient Valve (MCGV) Exchanging the Multi-Channel Gradient Valve (MCGV) NOTE The lifetime of the multi-channel gradient valve can be maintained by regularly flushing the valve, especially when using buffer solutions. If using buffer solutions, flush all channels of the valve with water to prevent precipitation of the buffer. Salt crystals can be forced into an unused channel and form plugs that may lead to leaks of that channel.
Repairing the Pump Exchanging the Multi-Channel Gradient Valve (MCGV) 1 Disconnect the connecting tube and the solvent tubes from the MCGV, unclip them from the tube clips and place them into the solvent cabinet to avoid flow by hydrostatic pressure. Unclip the waste tube from the cover and remove the waste funnel from the pump. 2 Press the lower sides of the cover to unclip it. Remove the cover.
Repairing the Pump Exchanging the Multi-Channel Gradient Valve (MCGV) 5 Replace the MCGV cover. Reconnect the waste funnel with the waste tube holder in the top cover. Insert waste tube in the holder in the waste pan and clip tube to the MCGV cover. Waste funnel Cover 6 Reconnect the tube from the active inlet valve to the middle position of the MCGV and then the solvent tubes at channel A to D of the MCGV.
Repairing the Pump Exchanging the optional Interface Board Exchanging the optional Interface Board CA UT IO N The interface board is sensitive to electrostatic discharge. Always use the ESD kit when handling electronic boards. When required Board defective Parts required BCD (Interface) board (G1351-68701), see “Optional Interface Boards” on page 216 1 Switch off the quaternary pump at the main power switch, unplug the pump from line power. 2 Disconnect cables from the interfaceboard connectors.
Repairing the Pump Exchanging Internal Parts Exchanging Internal Parts WAR NI N G The following procedures require opening the main cover of the quaternary pump. Always ensure the quaternary pump is disconnected from line power when the main cover is removed. The security lever at the power input socket prevents that the pump cover is taken off when line power is still connected. WAR NI N G To disconnect the quaternary pump from line, unplug the power cord.
Repairing the Pump Removing the Top Cover and Foam Removing the Top Cover and Foam Tools required Screwdriver Pozidriv #1 Preparations for this procedure Switch off quaternary pump at the main power switch. Disconnect the solvent inlet and outlet tubes from the pump. Beware of leaking solvents due to hydrostatic flow. Remove the solvent cabinet from the quaternary pump. Remove leak funnel with the waste tube from the top cover of the quaternary pump.
Repairing the Pump Removing the Top Cover and Foam 3 Lift the clips on both sides of the top cover (1). Remove the top cover (2). 4 Unscrew the screws on the top plate and remove the plate by lifting its back first and then sliding to the front. Screws (1) (1) (2) Cover Cover clip clip 5 If an optional interface board is installed, remove it from the pump slot. 6 Remove the top foam.
Repairing the Pump Removing the Top Cover and Foam Do not connect a power plug to the quaternary pump after removing the top covers. A safety light switch on the main board will prevent operation when the covers are removed. 7 Note the position of the safety light switch on the main board. Top foam cover The next figure shows the position of the safety light switch on the board.
Repairing the Pump Exchanging the Low Pressure Pump Main Board (LPM Board) Exchanging the Low Pressure Pump Main Board (LPM Board) When required Board defective Tools required Wrench 14 mm Wrench 7 mm Wrench 5 mm Parts required LPM Board, G1311-66520, exchange part number G1311-69520 1 Turn off the pump, disconnect all cables and remove the pump from the stack. 2 Remove the top cover and foam (see “Removing the Top Cover and Foam” on page 128). 3 Disconnect all connectors from the main board.
Repairing the Pump Exchanging the Low Pressure Pump Main Board (LPM Board) Table 8 Figure 21 Identifying Connectors on the LPM Board Connector Description Connector Description J1 RS232 J17 Encoder J2 Remote J19 AIV J3 Analog Pressure signal J23 Leak sensor J4 GPIB J24 Damper J5/6 CAN connector J25 Fan J9 Power supply J26 Interface board J16 Motor J30 MCGV Removing the Screws at the Back Plane.
Repairing the Pump Exchanging the Low Pressure Pump Main Board (LPM Board) WAR NI N G The RFI spring plate sitting on the board connectors is very sharp! Be careful, not to cut yourself, when removing it from the old board an sliding it onto the new board. 6 In most cases the RFI spring plate (Radio Frequency Interference spring plate, prevents radio emissions from the instrument to ambient) remains on the interface connectors of the board.
Repairing the Pump Exchanging the Low Pressure Pump Main Board (LPM Board) Entering the Type Command NOTE The main board is physically identical for the isocratic and the quaternary pump. After the installation of a new mainboard the TYPE (isocratic or quaternary) of the module is normally automatically detected. The specific TYPE tells the pump how to configure itself during turn on.
Repairing the Pump Exchanging the Low Pressure Pump Main Board (LPM Board) 8 Press the Execute (F8) key. Below the box, a reply line should then say: Reply RA 0000 TYPE "G1311A" . 9 Turn off the module, then turn it on again. Turn on should be normal. In the Records screen, the product# column should indicate the quaternary pump. If a chemstation is also connected, re-boot it now.
Repairing the Pump Exchanging the Low Pressure Pump Main Board (LPM Board) 4 Press Enter. This will display the box labeled Serial#. 5 Letters and numbers are created using the up and down arrows. Into the box labeled Serial#, enter the 10-character serial number for the quaternary pump. When the 10-character serial number is entered, press Enter to highlight the complete serial number. Then, press Done (F6). NOTE For firmware revisions below A02.
Repairing the Pump Exchanging the Low Pressure Pump Main Board (LPM Board) print sendmodule$ (lpmp, "ser?") The reply line will give the serial number of the quaternary pump. Replacing the Quaternary Pump’s Firmware The installation of new firmware is required • if a new version solves problems of the currently installed version. • if the version of firmware on the new main board (LPM) after an exchange of the board is older than the one previously installed.
Repairing the Pump Exchanging the Damper Exchanging the Damper When required No pressure output or when leaking Tools required Screwdriver Pozidriv #1 Wrench 1/4 inch Parts required Damper 79835-60005 1 Switch off the pump at the main power switch, remove all cables and tubings, remove the pump from the stack. Remove the front cover, top cover and top foam section (see “Removing the Top Cover and Foam” on page 128).
Repairing the Pump Exchanging the Damper 3 Disconnect the two damper capillaries. 4 Loosen the screws of the Z-panel and take it out of the instrument. Screws Damper capillaries 5 Disconnect the damper connector at the main board (J24). Lift the damper out of its position. Damper Damper Connector Z-panel 6 Place the new damper into the recess and connect to the main board (J24).
Repairing the Pump Exchanging the Damper 7 Place the Z-panel into position and fix it with the two screws. 8 Reconnect the two damper capillaries. Screws Z-panel 9 Clip the valve cover into its position and connect the tubings back to the valve ports. 140 Reconnect 10 Replace the top foam section, optional interface board (if installed), metal cover and top cover (see “Replacing the Top Cover and Foam” on page 156). Replace the pump on stack, reconnect all tubings and cables.
Repairing the Pump Exchanging the Fan Exchanging the Fan When required Fan not running Tools required Screwdriver Pozidriv #1 Parts required Fan, part number 3160-1016 1 Switch off the pump at the main power switch, remove all cables and tubings, remove the pump from the stack. 2 Remove the front cover, top covers and foam (see “Removing the Top Cover and Foam” on page 128) 3 Disconnect the fan connector at the processor board (J25).
Repairing the Pump Exchanging the Fan Figure 22 Exchanging the Fan J25, fan Air flow 142
Repairing the Pump Exchanging the Pump Drive Exchanging the Pump Drive WAR NI N G Never start the pump when the pump head is removed. This may damage the pump drive. When required: ❏ Error message: Motor Drive Power Tools required: ❏ Screwdriver Pozidriv #1 1 Switch off the pump at the main power switch, remove all cables and tubings, remove the pump from the stack. Remove the front cover, top cover and top foam section (see “Removing the Top Cover and Foam” on page 128).
Repairing the Pump Exchanging the Pump Drive 2 Remove the tubings from the gradient valve (MCGV) and remove the valve cover (see “Exchanging the Multi-Channel Gradient Valve (MCGV)” on page 123). 3 Disconnect all capillaries and tubes from the pump head and disconnect the active inlet valve connector. Cover 4 Loosen the screws of the Z-panel and take it out of the instrument. . 5 Using a 4-mm hexagonal key stepwise loosen and remove the two pump head screws and remove the pump head from the pump drive.
Repairing the Pump Exchanging the Pump Drive 6 Disconnect the pump drive cables from the main board (J16, J17) and lift the drive out of the foam. 7 Place the new pump drive into the recess in the foam part and connect the cables to the connectors on the main board (J16 Motor, J17 Encoder). J16 J16 J17 J17 8 Reinstall the pump head and fix with the two screws. 9 Replace the Z-panel and fix it with the two screws.
Repairing the Pump Exchanging the Pump Drive 10 Replace the MCGV cover and connect the connection tube to the middle position of the MCGV (see “Exchanging the Multi-Channel Gradient Valve (MCGV)” on page 123). Cover 146 11 Replace the pump on stack, reconnect all tubings and cables.
Repairing the Pump Exchanging the Power Supply Exchanging the Power Supply 1 Remove the pump from the stack, remove the front cover, top cover and top foam section (see “Removing the Top Cover and Foam” on page 128).
Repairing the Pump Exchanging the Power Supply 4 Disconnect the connectors of the pump assembly (J16, J17), fan (J25) and damper (J24) at the main board and lift the three assemblies out of the foam. Disconnect all remaining connectors at the main board. 5 Remove the LPM board, see “Exchanging the Low Pressure Pump Main Board (LPM Board)” on page 131. 6 Push the leak sensor cable through the recess in the foam and lift the bottom foam out of the instrument.
Repairing the Pump Exchanging the Power Supply 8 Unclip the power supply light pipe from the power supply and pull out the coupler. 9 Lift the power supply out of the unit. Coupler (2) (1) Light pipe 10 Place the new power supply into the instrument and fix the two screws at the back plane. 11 Install the coupler onto the power supply switch and clip the light pipe back onto the coupler.
Repairing the Pump Exchanging the Power Supply 12 Slide the leak sensor cable through the foam and replace the bottom foam. 13 Reinstall the main board, see “Exchanging the Low Pressure Pump Main Board (LPM Board)” on page 131. 14 Re-install pump assembly, fan and damper into the bottom foam and re-connect to the main board. 15 Place the Z-panel into its position and tighten the two screws.
Repairing the Pump Exchanging the Power Supply 16 Clip the MCGV cover into position. Reconnect all capillaries, tubes and cables to the pump head, the MCGV and the damper. 17 Install foam and top cover, see “Replacing the Top Cover and Foam” on page 156. Replace the pump on stack, reconnect all tubings and cables.
Repairing the Pump Exchanging the Leak Sensor Exchanging the Leak Sensor When required Leak messages without leak in the leak pan Tools required Screwdriver Pozidriv #1 Wrench 1/4 inch Wrench 14 mm Wrench 7 mm Wrench 5 mm Parts required Leak sensor 5061-3356 1 Switch off the pump at the main power switch, remove all cables and tubings, remove the pump from the stack. 2 Remove the top covers and foam, see “Removing the Top Cover and Foam” on page 128.
Repairing the Pump Exchanging the Leak Sensor 13 Replace the pump on stack, reconnect all tubings and cables.
Repairing the Pump Exchanging the Status Light Pipe Exchanging the Status Light Pipe When required If part is broken Tools required Screwdriver Pozidriv #1 Part required Status light pipe 5041-8384 Preparation for this procedure: ❏Remove the front cover and top cover, see “Removing the Top Cover and Foam” on page 128. 1 The status light pipe is clipped into the top cover. 154 2 Replace the top cover, see “Replacing the Top Cover and Foam” on page 156.
Repairing the Pump Assembling the Main Cover Assembling the Main Cover When required If cover is broken Tools required None Parts required Cover kit 5062-8565 (includes base, top, left and right) NOTE The cover kit contains all parts, but it is not assembled. WAR NI N G In case you insert the left or right side in the opposite position, you may not be able to remove the side from the top part. 1 Place the top part on the bench and insert the left and right side into the top part.
Repairing the Pump Replacing the Top Cover and Foam Replacing the Top Cover and Foam Tools required Screwdriver Pozidriv #1 Preparations for this procedure Make sure that after your repair all assemblies, cables, capillaries and connectors are located in its correct place. 1 Replace the top foam section. 2 Make sure that the foam is installed correctly and is located in the safety light switch.
Repairing the Pump Replacing the Top Cover and Foam 3 Location of foam in the light switch. 4 Replace the optional interface board. Light switch 5 Replace the metal cover (slide the metal tabs into place (1) underneath the Z-Panel in the front, then lower the back of the metal plate (2)) and fix the two holding screws. 6 Replace the top cover. Replace the pump on stack, reconnect all tubings and cables.
Repairing the Pump Replacing the Top Cover and Foam 7 Ensure clips are seated correctly and, move the lever back. 8 Place the quaternary pump back to its position in the stack, place vacuum degasser and solvent cabinet back in place and re-connect all tubes and capillaries. Lever 9 Replace the leak funnel with the waste tube. Locate the lower end of the waste tube in the holder of the leak pan and clip tube to the holder at the gradient valve cover.
5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials
Overview of Main Assemblies Figure 24 Overview of Main Assemblies (Front View) 5 4 3 2 1 6 7 11 8 10 9 160
Parts and Materials Overview of Main Assemblies Table 9 Repair Parts - Pump Housing and Main Assemblies (Front View) Item Description Part Number 1 Pump head, see page 172 G1311-60004 2 Pump drive assembly G1311-60001 Exchange assembly — pump drive G1311-69001 3 Cable assembly — AIV to main board G1311-61601 4 Low-pressure pump main board (LPM) G1311-66520 Exchange assembly — LPM board G1311-69520 5 Cable to MCGV G1311-61600 6 Fan assembly 3160-1016 7 Damping unit 79835-60005 8
Parts and Materials Overview of Main Assemblies Figure 25 Overview of Main Assemblies (Rear View) (3) (1) (2) (4) (5) (6) (7) Table 10 Repair Parts - Pump Housing and Main Assemblies (Rear View) Item Description Part Number 1 Nut M14 — analog output 2940-0256 2 Washer — analog output 2190-0699 3 Standoff — remote connector 1251-7788 4 Standoff — GPIB connector 0380-0643 5 Power supply (behind rear panel) 0950-2528 6 Screw, M4, 7 mm lg — power supply 0515-0910 7 Washer — power su
Parts and Materials Control Module (B-version) Control Module (B-version) Table 11 Control Module (B-version) Item Figure 26 Description Part Number Control module, replacement part including cable G1323-67001 Plastic Housing Kit, includes front, back and a clamp 5062-8583 CAN cable, Agilent 1100 module to control module G1323-81600 Control Module B-version 163
Parts and Materials Solvent Cabinet Solvent Cabinet Table 12 Figure 27 Solvent Cabinet Parts Item Description Part Number 1 Solvent cabinet, including all plastic parts 5062-8581 2 Name plate, Agilent 1100 5042-1312 3 Front panel, solvent cabinet 5062-8580 4 Leak pan, solvent cabinet 5042-1307 5 Bottle transparent 9301-1420 6 Bottle amber 9301-1450 7 Bottle-head assembly, see page 165 G1311-60003 Solvent Cabinet Parts 7 2 1 3 164 4 5/6
Parts and Materials Bottle Head Assembly Bottle Head Assembly Table 13 Bottle-Head Assembly Parts Item Figure 28 Description Part Number Complete assembly G1311-60003 1 Ferrules with lock ring 5063-6598 (10x) 2 Tube screw 5063-6599 (10x) 3 Wire marker No part number 4 Solvent tubing, 5 m 5062-2483 5 Frit adapter (pack of 4) 5062-8517 6 Solvent inlet filter 5042-1347 Bottle-Head Assembly Parts 3 4 2 1 5 6 165
Parts and Materials Hydraulic Path Hydraulic Path Table 14 Figure 29 Hydraulic Path Item Description Part Number 1 Outlet capillary, pump to injector device G1312-67305 Outlet Capillary, pump to thermostattable autosampler G1329-87300 Bottle-head assembly, bottle to vacuum degasser (see “Bottle Head Assembly” on page 165) G1311-60003 2 Solvent tube, vacuum degasser to MCGV (pack of 4) G1322-67300 3 Capillary, plunger 1 to damper G1311-67301 4 Capillary, damper to plunger 2 G1311-67300
Parts and Materials Cover Parts Cover Parts Table 15 Figure 30 Cover Parts Item Description Part Number 1 Cover kit (includes top, both sides, base) 5062-8565 2 Front plate 5062-8566 3 Logo plate, Agilent 1100 5042-1312 Cover Parts 1 (top) 3 1 (base) 1 (side) 2 167
Parts and Materials Sheet Metal Kit Sheet Metal Kit Table 16 Figure 31 Sheet Metal Kit Item Description Part Number 1 Sheet metal kit, includes top, base and Z-panel G1311-68701 2 Screw, for cover and Z-Panel 5022-2112 3 Board cover 5001-3772 Sheet Metal Kit 2 3 2 1 Z-panel 168
Parts and Materials Foam Parts Foam Parts Table 17 Figure 32 Foam Parts Item Description Part Number 1 Foam Kit (includes upper and lower foam) G1311-68702 2 Bushing, for pump drive 1520-0404 3 Board Guides 5041-8395 Foam Parts 3 3 1 (upper foam) 2 2 1 (lower foam) 169
Parts and Materials Power and Status Light Pipes Power and Status Light Pipes Table 18 Figure 33 Power and Status Light Pipes Item Description Part Number 1 Power switch coupler 5041-8383 2 Light pipe — power switch 5041-8382 3 Power switch button 5041-8381 4 Light pipe — status lamp 5041-8384 Power and Status Light Pipes 1 2 4 170 3
Parts and Materials Leak Parts Leak Parts Table 19 Figure 34 Leak Parts Item Description Part Number 1 Leak sensor 5061-3356 2 Tube clip 5041-8387 3 Leak pan, pump 5041-8390 4 Leak funnel 5041-8388 5 Holder, leak funnel 5041-8389 6 Corrugated waste tube (reorder pack), 5m 5062-2463 Leak Parts 1 2 3 4 5 6 171
Parts and Materials Pump Head Assembly Pump Head Assembly Table 20 Pump Head Assembly Item Description Part Number Complete assembly, including items marked with * G1311-60004 1* Sapphire plunger 5063-6586 2* Plunger housing (including spring) G1311-60002 3* Support ring 5001-3739 4* Seal (pack of 2) or Seal (pack of 2), for normal phase applications 5063-6589 0905-1420 5* Pump chamber housing G1311-25200 6 Active inlet valve (complete with cartridge) G1312-60010 Replacement cartri
Parts and Materials Pump Head Assembly Figure 35 Pump Head Assembly 10 7 9 10 8 5 4 2 3 6 1 173
Parts and Materials Pump Head Assembly with Seal Wash Option Pump Head Assembly with Seal Wash Option Table 21 Pump Head Assembly with Seal Wash Option Item Description Part Number Complete assembly, including parts marked with * G1311-60005 1* Sapphire plunger 5063-6586 2* Plunger housing (including spring) G1311-60002 3* Support ring, seal wash 5062-2465 4* Secondary seal 0905-1175 5* Wash tube (1.
Parts and Materials Pump Head Assembly with Seal Wash Option Figure 36 Pump Head with Seal Wash Option 14 11 13 12 14 9 8 5 4 6 7 10 2 3 1 175
Parts and Materials Outlet Ball Valve Assembly Outlet Ball Valve Assembly Table 22 Outlet Ball Valve Assembly Item Figure 37 Description Part Number Outlet ball valve — complete assembly G1311-60012 1 Socket cap 01018-25209 2 Outlet valve housing screw 01018-22410 3 Gold seal, outlet 5001-3707 4 Cap (pack of 4) 5062-2485 Outlet Ball Valve Assembly 1 2 3 4 176
Parts and Materials Purge Valve Assembly Purge Valve Assembly Table 23 Purge-Valve Assembly Item Figure 38 Description Part Number Purge valve — complete assembly G1311-60009 1 Valve body No part number 2 PTFE frit (pack of 5) 01018-22707 3 Gold seal 5001-3707 4 Cap (pack of 4) 5062-2485 Purge-Valve Assembly 1 2 3 4 177
Parts and Materials Active Inlet Valve Assembly Active Inlet Valve Assembly Table 24 Active Inlet Valve Assembly Item Figure 39 Description Part Number Active inlet valve — complete assembly G1312-60010 1 Valve body No part number 2 Valve cartridge 5062-8562 Active Inlet Valve Assembly Valve body 1 Valve cartridge 178
Parts and Materials Accessory Kit G1311-68705 Accessory Kit G1311-68705 Table 25 Tools and Accessories Description Part Number Wrench 14 mm 8710-1924 Seal insert tool 01018-23702 PTFE Frit (pack of 5) 01018-22707 Corrugated waste tube (1.2 m) no PN Corrugated waste tube (reorder number), 5m 5062-2463 Velocity regulator (reorder number, 3) 5062-2486 ESD wrist strap 9300-1408 Hex key 4mm 8710-2392 Wrench 1/4 – 5/16 inch 8710-0510 Capillary, pump to injection device, 600 mm lg, ID 0.
Seal Wash Option Kit 01018-68722 Table 26 Seal Wash Option Description Part Number Support ring, seal wash (pack of 2) 5062-2465 Secondary seal (pre-installed in support ring) 0905-1175 Seal keeper (pack of 2) 5001-3743 Wash tube (1m) 0890-1764 Velocity regulator* 5062-2486 Seals insert tool 01018-23702 Seal (pack of 2) 5063-6589 Syringe ** 5062-8534 Syringe adapter 0100-1681 Knife no PN Sanding Paper no PN * ** Reorder number (pack of 3) Reorder number (pack of 10) 180
Parts and Materials Cable Overview Cable Overview WAR NI N G Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
Parts and Materials Cable Overview Table 27 Cables Overview, continued Type BCD cables Description Part Number HP 1040 diode-array detector 01046-60202 HP 1090 liquid chromatographs 01046-60202 Signal distribution module 01046-60202 3392/3 integrators 18594-60510 3396 integrator 03396-60560 General purpose (spade Lugs) 18594-60520 Auxiliary Agilent 1100 Series vacuum degasser G1322-61600 CAN cables Agilent 1100 module to module,0.
Parts and Materials Analog Cables Analog Cables One end of these cables provides a BNC connector to be connected to Agilent 1100 Series modules. The other end depends on the instrument to which connection is being made.
Parts and Materials Analog Cables Agilent 1100 to BNC Connector Connector 8120-1840 Pin BNC Pin Agilent 1100 Signal Name Shield Shield Analog - Center Center Analog + Pin Agilent 1100 Signal Name Agilent 1100 to General Purpose Connector 01046-60105 Pin 3394/6 1 184 Not connected 2 Black Analog - 3 Red Analog +
Parts and Materials Remote Cables Remote Cables One end of these cables provides a Agilent Technologies APG (Analytical Products Group) remote connector to be connected to Agilent 1100 Series modules. The other end depends on the instrument to be connected to.
Parts and Materials Remote Cables Agilent 1100 to 3392/3 Integrators Connector 01046-60206 4 - Key Pin 3392/3 Pin Agilent 1100 Signal Name Active (TTL) 3 1 - White Digital ground NC 2 - Brown Prepare run Low 11 3 - Gray Start Low NC 4 - Blue Shut down Low NC 5 - Pink Not connected NC 6 - Yellow Power on High 9 7 - Red Ready High 1 8 - Green Stop Low NC 9 - Black Start request Low Pin 3394 Pin Agilent 1100 Signal Name Active (TTL) 9 1 - White Digital ground NC
Parts and Materials Remote Cables Agilent 1100 to 3396A Integrators Connector 03394-60600 Pin 3394 Pin Agilent 1100 Signal Name 9 1 - White Digital ground NC 2 - Brown Prepare run Low 3 3 - Gray Start Low NC 4 - Blue Shut down Low NC 5 - Pink Not connected NC 6 - Yellow Power on High 5,14 7 - Red Ready High 1 8 - Green Stop Low NC 9 - Black Start request Low 13, 15 Active (TTL) Not connected Agilent 1100 to 3396 Series II / 3395A Integrators Use the cable 03394-6060
Parts and Materials Remote Cables Agilent 1100 to 3396 Series III / 3395B Integrators Connector 03396-61010 Pin 33XX Pin Agilent 1100 Signal Name 9 1 - White Digital ground NC 2 - Brown Prepare run Low 3 3 - Gray Start Low NC 4 - Blue Shut down Low NC 5 - Pink Not connected NC 6 - Yellow Power on High 14 7 - Red Ready High 4 8 - Green Stop Low NC 9 - Black Start request Low 13, 15 Active (TTL) Not connected Agilent 1100 to HP 1050, HP 1046A or Agilent 35900 A/D Conv
Parts and Materials Remote Cables Agilent 1100 to HP 1090 LC, HP 1040 DAD or Signal Distribution Module Connector 01046-60202 5 - Key Pin HP 1090 Pin Active Agilent 1100 Signal Name (TTL) 1 1 - White Digital ground NC 2 - Brown Prepare run Low 4 3 - Gray Start Low 7 4 - Blue Shut down Low 8 5 - Pink Not connected NC 6 - Yellow Power on High 3 7 - Red Ready High 6 8 - Green Stop Low NC 9 - Black Start request Low Pin Agilent 1100 Signal Name Active (TTL) 1 - White D
Parts and Materials BCD Cables BCD Cables One end of these cables provides a 15-pin BCD connector to be connected to the Agilent 1100 Series modules. The other end depends on the instrument to be connected to.
Parts and Materials BCD Cables Agilent 1100 to 3396 Integrators Connector 03396-60560 Pin 3392/3 Pin Agilent 1100 Signal Name BCD Digit 1 1 BCD 5 20 2 2 BCD 7 80 3 3 BCD 6 40 4 4 BCD 4 10 5 5 BCD 0\ 1 6 6 BCD 3 8 7 7 BCD 2 4 8 8 BCD 1 2 9 9 Digital ground NC 15 +5V Low Wire Color Pin Agilent 1100 Signal Name BCD Digit Green 1 BCD 5 20 Violet 2 BCD 7 80 Blue 3 BCD 6 40 Yellow 4 BCD 4 10 Black 5 BCD 0\ 1 Orange 6 BCD 3 8 Red 7 BCD 2 4
Parts and Materials Auxiliary Cable Auxiliary Cable One end of this cable provides a modular plug to be connected to the Agilent 1100 Series vacuum degasser. The other end is for general purpose.
Parts and Materials External Contact Cable 5 10 15 External Contact Cable 1 6 11 One end of this cable provides a 15-pin plug to be connected to Agilent 1100 Series module’s interface board. The other end is for general purpose.
Parts and Materials RS-232 Cable Kit RS-232 Cable Kit This kit contains a 9-pin female to 9-pin female Null Modem (printer) cable and one adapter. Use the cable and adapter to connect Agilent Technologies instruments with 9-pin male RS-232 connectors to most PCs or printers.
Parts and Materials LAN Cables LAN Cables Recommended Cables For point to point connection (not using a network hub) use a twisted pair cross over LAN cable (P/N 5183-4649, 10 feet long). For standard network connections using a hub use category 5 UTP cables, (P/N G1530-61480, 8 m long).
Parts and Materials LAN Cables 196
6 6 Introduction to the Quaternary Pump An introduction to the pump, instrument overview, theory of operation, external communication and internal connectors
Introduction to the Quaternary Pump Introduction to the Quaternary Pump Introduction to the Quaternary Pump The quaternary pump comprises a solvent cabinet, a vacuum degasser and a four-channel gradient pump. The four-channel gradient pump comprises a high-speed proportioning valve and a pump assembly. It provides gradient generation by low pressure mixing. Degassing is a must for a low-pressure gradient system therefore the Agilent 1100 Series vacuum degasser is part of the quaternary pump system.
Introduction to the Quaternary Pump Overview Overview The quaternary pump is based on a two-channel, dual plunger in-series design which comprises all essential functions that a solvent delivery system has to fulfill. Metering of solvent and delivery to the high-pressure side are performed by one pump assembly which can generate pressure up to 400 bar. Degassing of the solvents is done in a vacuum degasser and solvent compositions are generated on the low-pressure side by a high-speed proportioning valve.
Introduction to the Quaternary Pump Overview How does the Pump Work? The liquid runs from the solvent reservoir through the degasser to the MCGV and from there to the active inlet valve. The pump assembly comprises two substantially identical plunger/chamber units. Both plunger/chamber units comprise a ball-screw drive and a pump head with one sapphire plunger for reciprocating movement in it. A servo-controlled variable reluctance motor drives the two ball-screw drives in opposite directions.
Introduction to the Quaternary Pump Overview Figure 42 Principle of the Quaternary Pump Damper Chamber 2 Chamber 1 Purge valve To mixing chamber Inlet valve Outlet valve To waste From solvent bottle Seal Plunger 1 Plunger 2 Ball screw drive Gear Motor with encoder When turned on, the quaternary pump runs through an initialization procedure to determine the upper dead center of the first plunger.
Introduction to the Quaternary Pump Overview chamber is pressed through the outlet ball valve into the second chamber. The second plunger draws in half of the volume displaced by the first plunger and the remaining half volume is directly delivered into the system. During the drawing stroke of the first plunger, the second plunger delivers the drawn volume into the system.
Introduction to the Quaternary Pump Overview With a compressibility value set the processor calculates a compensation volume, that is depending on the backpressure in the system and the selected compressibility. This compensation volume will be added to the normal stroke volume and compensates the previous described loss of volume during the delivery stroke of the first plunger.
Introduction to the Quaternary Pump Electrical Connections Electrical Connections • The GPIB connector is used to connect the pump with a computer. The address and control switch module next to the GPIB connector determines the GPIB address of your pump. The switches are preset to a default address (Table 31 on page 218 or Table 35 on page 223). This address is recognized at powercycling the module. • The CAN bus is a serial bus with high speed data transfer.
Introduction to the Quaternary Pump Electrical Connections Figure 43 Electrical Connections Security lever Slot for interface board Analog pressure, 2mV/bar APG Remote RS-232C CAN GPIB Power Configuration switch WAR NI N G To disconnect the quaternary pump from line, unplug the power cord. The power supply still uses some power, even if the power switch on the front panel is turned off.
Introduction to the Quaternary Pump Instrument Layout Instrument Layout The industrial design of the module incorporates several innovative features. It uses Agilent’s E-PAC concept for the packaging of electronics and mechanical assemblies. This concept is based upon the use of expanded polypropylene (EPP) layers foam plastic spacers in which the mechanical and electronic boards components of the module are placed.
Introduction to the Quaternary Pump Early Maintenance Feedback (EMF) Early Maintenance Feedback (EMF) Maintenance requires the exchange of components in the flow path which are subject to mechanical wear or stress. Ideally, the frequency at which components are exchanged should be based on the intensity of usage of the instrument and the analytical conditions, and not on a predefined time interval.
Introduction to the Quaternary Pump Early Maintenance Feedback (EMF) Using the EMF Counters The user-settable EMF limits for the EMF counters enable the early maintenance feedback to be adapted to specific user requirements. The wear of pump components is dependent on the analytical conditions, therefore, the definition of the maximum limits need to be determined based on the specific operating conditions of the instrument.
Introduction to the Quaternary Pump The Electronics The Electronics The electronics are comprised of four main components: • The low pressure pump main board (LPM), see page 210. • Power supply, see page 228. Optional: • Interface board (BCD/external contacts), see page 216. • Interface board (LAN), see page 217.
Introduction to the Quaternary Pump The Low-Pressure Pump Main Board (LPM) The Low-Pressure Pump Main Board (LPM) The board controls all information and activities of all assemblies within the quaternary pump. The operator enters parameters, changes modes and controls the quaternary pump through interfaces (CAN, GPIB or RS-232C) connected to the user-interfaces. Figure 45 and Figure 46 show the block diagrams of this board.
Introduction to the Quaternary Pump The Low-Pressure Pump Main Board (LPM) Fan Drive The fan speed is controlled by the main processor depending on the internal heat distribution in the quaternary pump. The fan provides a PWM signal which is proportional to its speed. This fan status signal is used for diagnostics. Electronic Fuses The valve circuits are electronically fused on the board.
Introduction to the Quaternary Pump The Low-Pressure Pump Main Board (LPM) Figure 45 Block Diagram Low Pressure Main Board (LPM) MCGV Active inlet valve 4 valve drives ASIC Valve drive Current control Pump head Motor Motor control 1 Data/ address Drive ϑ Control Encoder Processor Motor drive Multiplexer Damper Pressure sensor Filter & amplifier A/D converter 1 Integrator Analog output Offset correction pressure converter Leak sensor Amplifier & ϑ ambient measurement Fan Drive Senso
Introduction to the Quaternary Pump The Low-Pressure Pump Main Board (LPM) Figure 46 Interconnection Diagram Low Pressure Main Board (LPM) Motor Hardware switch J16 Motor drive Encoder AIV Z-panel MCGV Z-panel J17 +36 V +24 V +/-15V Processor J19 Inlet valve drive +36V Electronic fuse J30 MCGV drive C&D +36 V Electronic fuse MCGV drive A&B +36 V Hardware switch +24 V Electronic fuse +15 V Leak converter +24 V Fan drive J23 Leak sensor J25 Fan J24 +/-15V +36 V +24 V +/-15V
Introduction to the Quaternary Pump Firmware Description Firmware Description The firmware of the instrument consists of two independent sections: • a non-instrument specific section, called ‘resident system’, • an instrument specific section, called ‘main system’. Resident System This resident section of the firmware is identical for all Agilent 1100 series modules.
Introduction to the Quaternary Pump Firmware Description Firmware Updates Firmware updates can be done using your user interface: • handheld control module with files from a PC-card or • Agilent ChemStation with files from floppy disk The file naming conventions are: xxxx-vvv.DLB, where xxxxis the product number, e.g. 1311 for the G1311A QuaternaryPump), and vvvis the revision number, for example 200 is revision 2.00. For instructions refer to your user interface.
Introduction to the Quaternary Pump Optional Interface Boards Optional Interface Boards The Agilent 1100 Series modules have one optional board slot that allows to add an interface board to the modules. Table 29 Optional Interface Boards Description Part Number BCD Board G1351-68701 Fuse 250 mA (four are on the board) 2110-0004 LAN Board (see next page for details) BCD Board The BCD board provides a BCD output for the bottle number of the Agilent 1100 Series autosampler and four external contacts.
Introduction to the Quaternary Pump Optional Interface Boards LAN Board The HP LAN board is actually an HP JetDirect card, which is a network interface card used in HP printers. NOTE One board is required per Agilent 1100 stack. If the Agilent 1100 stack has a DAD, then the DAD MUST be the module used for LAN board installation. If no DAD is present a pump should be used for the LAN board installation.
Introduction to the Quaternary Pump Agilent 1100 Series Interfaces Agilent 1100 Series Interfaces The Agilent 1100 Series modules provide the following interfaces: Table 31 Agilent 1100 Series Interfaces Interface Type Pumps Autosampler DA Detector MW Detector FL Detector CAN Yes Yes Yes Yes Yes No GPIB Yes Yes Yes Yes Yes No RS-232C Yes Yes Yes Yes Yes No APG Remote Yes Yes Yes Yes Yes Yes Analog Yes No 2× 1× No Yes* Interface board Yes Yes Yes Yes No No *
Introduction to the Quaternary Pump Agilent 1100 Series Interfaces WAR NI N G Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations, see “Cable Overview” on page 181. Analog Signal Output The analog signal output can be distributed to a recording device. For details refer to the description of the main board of the module. GPIB Interface The GPIB connector is used to connect the module with a computer.
Introduction to the Quaternary Pump Agilent 1100 Series Interfaces APG Remote Interface The APG Remote connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features as common shut down, prepare, and so on. Remote control allows easy connection between single instruments or systems to ensure coordinated analysis with simple coupling requirements. The subminiature D connector is used.
Introduction to the Quaternary Pump Agilent 1100 Series Interfaces Table 33 APG Remote Signal Distribution Pin Signal Description 1 DGND Digital ground 2 PREPARE (L) Request to prepare for analysis (for example, detector lamp on). Receiver is any module performing preanalysis activities. 3 START (L) Request to start run / timetable. Receiver is any module performing run-time controlled activities. 4 SHUT DOWN (L) System has serious problem (for example, leak: stops pump).
Introduction to the Quaternary Pump Agilent 1100 Series Interfaces connector. The RS-232C is designed as DCE (data communication equipment) with a 9-pin male SUB-D type connector.
Introduction to the Quaternary Pump Setting the 8-bit Configuration Switch Setting the 8-bit Configuration Switch The 8-bit configuration switch is located next to the GPIB connector. Switch settings provide configuration parameters for GPIB address, serial communication protocol and instrument specific initialization procedures. Figure 50 8-bit Configuration Switch 1 factory setting is shown for quaternary pump 0 .
Introduction to the Quaternary Pump Setting the 8-bit Configuration Switch In the non-volatile memory the parameters are kept, regardless of whether you turn the instrument off and on again. They will be kept until the same set of parameters is subsequently changed and power is reset. All other previously stored configuration settings will still remain in the non-volatile memory.
Introduction to the Quaternary Pump Setting the 8-bit Configuration Switch Switches 1 in down and 2 in up position define that the RS-232C parameters will be changed. Once the change has been completed, the instrument must be powered up again in order to store the values in the non-volatile memory.
Introduction to the Quaternary Pump Setting the 8-bit Configuration Switch Table 40 Parity Settings Switches Parity 7 8 0 0 No Parity 0 1 Odd Parity 1 0 Even Parity Forced Cold-Start Settings Switches 1 and 2 do not force storage of this set of parameters in non-volatile memory. Returning the switches 1 and 2 to other positions (other than being both up) will allow for normal operation. CA UT IO N A forced cold start erases all methods and data stored in non-volatile memory.
Introduction to the Quaternary Pump Setting the 8-bit Configuration Switch Switches 1 and 2 do not force storage of this set of parameters in non-volatile memory. Returning the switches 1 and 2 to other positions (other than being both up) will allow for normal operation. If you use the following switch settings and power the instrument up again, the instrument firmware stays in the resident part, that is, it is not operable as a quaternary pump.
Introduction to the Quaternary Pump The Main Power Supply Assembly The Main Power Supply Assembly The main power supply comprises a closed assembly (no component-level repair possibility). The power supply provides all DC voltages used in the quaternary pump. The line voltage can vary in a range from 100 – 120 or 220 – 240 volts AC ± 10 % and needs no manual setting.
Introduction to the Quaternary Pump The Main Power Supply Assembly When overload conditions occur, the power supply turns off all output voltages. Turning the line power off and on again resets the power supply to normal operation if the cause of the overload condition has been removed. An over-temperature sensor in the main power supply is used to turn off output voltages if the temperature exceeds the acceptable limit (for example, if the cooling fan of the quaternary pump fails).
Introduction to the Quaternary Pump The Main Power Supply Assembly 230
7 7 Control Module Screens for the Quaternary Pump
Control Module Screens for the Quaternary Pump This chapter is intended to introduce an operator to the screens available for operation of the Agilent 1100 quaternary pump with the Agilent 1100 control module. Please use the manual of the control module for further detailed reference.
Control Module Screens for the Quaternary Pump Screens available from the Analysis screen Screens available from the Analysis screen The Analysis screen This is the wake-up screen, if the Agilent 1100 quaternary pump is the only configured Agilent 1100 module. It is used to enter the most common pump method parameters. The m-key allows access to the context sensitive menu. Date&Time allows you to change time settings. Print Screen gives acces to the print configuration screen.
Control Module Screens for the Quaternary Pump Screens available from the Analysis screen Setup View In the Setup view, modules can be added or removed to the view. Here, e.g. the autosampler and thermostatted column compartement parameters are shown on the display as well. The number of parameters on the display is restricted as additional modules are added. A maximum of 4 modules is shown automatically. If more modules are connected to the system, you have to choose 4 of them in Setup view.
Control Module Screens for the Quaternary Pump Screens available from the Analysis screen Pump ON/OFF From the Analysis screen use the F7 key to proceed to the turn on screen. Press F8 (On) once to turn on the pump. If more than one module is available, select the quaternary pump from the pop-up menu. Settings With the Settings key you open a pull-down menu where you can select the quaternary pump module.
Control Module Screens for the Quaternary Pump Screens available from the Analysis screen Settings Within the Settings you can change the pump parameters. You have access to a different set of parameters available through the F1-5 keys. F7 key resets the pump to default values. F8 opens a window to turn on the pump. Use the m-key for the context sensitive menu. The Status command pulls up a monitor screen displaying signals and spectra as programmed. Reset will load the pump default parameters.
Control Module Screens for the Quaternary Pump Screens available from the Analysis screen Settings - Timetable With the F2 key (Timetable) you can list the timetable for the pump. Press F7 key (Insert) to add entries or F6 key (Delete) to remove entries. Use the F6 key (Done) to view the entered lines of the timetable. Use the m-key for the context sensitive menu. It gives you additional tools for the timetable.
Control Module Screens for the Quaternary Pump Screens available from the Analysis screen Settings - Pressure With the F3 key (Pressure) you can change the settings for the pressure limits. Settings - Bottle Fillings With the F4 key (Bottle Fillings) you can adjust the settings for the bottle fillings to their current state. Settings - Runtimes With the F5 key (Runtimes) you can change the stop time and the post-run time.
Control Module Screens for the Quaternary Pump Screens available from the Analysis screen Analog -Status Press F5 key (Views) and select Status. Status This is an example if an Agilent 1100 pump is configured standalone. Information on the actual flow rate , mobile phase composition, pressure and %-ripple, elapsed run time and the pressure plot are shown. Press key F8 (Start) to start a run, key F7 (Rescale) to maximize the signal.
Control Module Screens for the Quaternary Pump Screens available from the Analysis screen Press F6 key (Select). Here you can add additional online signals (maximum are 3). Additional signals could also be chromatograms or temperature signals from other modules. Use the Right/Left arrows to switch between Available and Selected Signals. Use the F8 key (Move) to enter available signals into the box for selected signals or vice versa.
Control Module Screens for the Quaternary Pump Screens available from the Analysis screen Use F2 key (PC-Card) to save a method on a PCMCIA card. Use the Right/Left arrows to switch between PC-Card and Instrument window. Use the UP/Down arrows to select the method. Use the F7/F8 keys (Copy) to enter available signals into the box for selected signals or vice versa.
Control Module Screens for the Quaternary Pump Screens available from the System screen Screens available from the System screen System screen Use the Esc key to receive Views on the F5 key. Choose System from the pull-down menu. This screen shows the last activities in the system. System - Control Use the F1 key (Control) to select the quaternary pump. Here you receive information about the not-ready conditions if needed. F2 key (Reset) does a re-initialization of the pump.
Control Module Screens for the Quaternary Pump Screens available from the System screen System Configuration On the System screen use the F2 key (Configure) to select the pump. Use the F1 key (Interfaces) to access the interface settings (if required). Press F4 (Bottle fillings) to adjust bottle fillings to the currrent state.
Control Module Screens for the Quaternary Pump Screens available from the Records screen Screens available from the Records screen Records screen Use the Esc key to receive Views on the F5 key. Choose System from the pull-down menu. Use the F4 key (Records) to select the pump. Errors are reported either into the System Log (F2) or Error Log (F3). System / Error Log Use the F2 key (System Log) or F3 key (Error Log) to look for errors.
Control Module Screens for the Quaternary Pump Screens available from the Records screen Info Log Use the m-key to receive a pop-up menu, Select Info Log. A list of the last events are listed. For troubleshooting reasons they can be printed or saved to a file on the PCMCIA card (using the m-key for the context sensitive menu). EMF (Early Maintenance Feedback) Use the F1 key (EMF) to set EMF parameters.
Control Module Screens for the Quaternary Pump Screens available from the Records screen If a set limit has been exceeded, a message box will pop up. This will not stop a sequence or run (information only to plan maintenance activities). If you press Reset, the limits will be removed. Ignore will continue to keep the EMF flag set. Firmware Update Use the Esc key to receive Views on the F5 key. Choose System from the pull-down menu. Use the F3 key (Records) to select the pump.
Control Module Screens for the Quaternary Pump Screens available from the Records screen Use the Esc key to receive Views on the F5 key. Choose System from the pull-down menu. Use the F3 key (Records) to select the Generic module. In this screen the resident firmware revision is shown. Use the F5 key (FW Update) to enter the Update section. Select the a file from the PCMCIA card (1311nnnn.DLB) and press execute.
Control Module Screens for the Quaternary Pump Screens available from the Records screen Changing the serial number In case the serial number of the module has to be added, use the m-key to open the menu Enter Serial#. The serial number becomes active after restart of the module. Maintenance activities On the Records screen use the F4 key (Maint log) to view and edit the maintenance logbook.
Control Module Screens for the Quaternary Pump Screens available from the Records screen Use the F7 key (Add) to add new maintenance activities. If an activity is not listed, you can type the activity into the line “Add” using the control modules key pad. Changting the Type In order to change the type of the module (this may be necessary after an exchange of the mainbord) press ’m.m’ in the Test-screen and select Command.
Control Module Screens for the Quaternary Pump Screens available from the Records screen Enter ’TYPE G1311A’ in the Instr-line in order to configure the module as a quaternary pump. Schematics Select Schematics after pressing ’m.m’ on the Test-screen in order to get a schematic overview of the quaternary pump.
Control Module Screens for the Quaternary Pump Diagnostics and Tests Diagnostics and Tests Tests screen Use the Esc key to receive Views on the F5 key. Choose System from the pull-down menu. Use the F3 key (Tests) to select the pump. Two tests are available to test the Agilent 1100 pumps. Leak Test Press F1 (Leak Test) on the Test screen to perform a leak test.
Control Module Screens for the Quaternary Pump Diagnostics and Tests Pressure Test Use the F2 key (Pressure Test) to perform a pressure test of the system. Several steps like purging the system, setting up Isopropanol as solvent on channel D and blocking the column outlet with a blank nut have to be performed before operating the pressure test. For details use the ’i’-key to achieve context sensitive help, follow the instructions on the screen and refer to “Running the Pressure Test” on page 84.
8 8 Specifications Performance specifications of the quaternary pump
Specifications Performance Specifications Performance Specifications Table 44 Performance Specification Agilent 1100 Series Quaternary Pump Type Specification Hydraulic system Dual plunger in series pump with proprietary servo-controlled variable stroke drive, floating plungers and active inlet valve Setable flow range 0.001 – 10 ml/min, in 0.001 ml/min increments Flow range 0.2 – 10.0 ml/min Flow precision < 0.3 % RSD (typically 0.
Specifications Performance Specifications Table 44 Performance Specification Agilent 1100 Series Quaternary Pump Analog output For pressure monitoring, 2 mV/bar, one output Communications Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional Safety and maintenance Extensive diagnostics, error detection and display (through control module and Agilent ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping s
Specifications Performance Specifications 256
Warranty Statement Warranty Statement All Chemical Analysis Products Agilent Technologies warrants its chemical analysis products against defects in materials and workmanship. For details of the warranty period in your country, call Agilent. During the warranty period, Agilent will, at its option, repair or replace products which prove to be defective. Products that are installed by Agilent are warranted from the installation date, all others from the ship date.
Warranty Statement 4 operation outside of the environmental and electrical specifications for the product, 5 improper site preparation and maintenance, or 6 customer induced contamination or leaks. THE WARRANTY SET FORTH IS EXCLUSIVE AND NO OTHER WARRANTY, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED. AGILENT SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Warranty Statement Responsibilities of Agilent Technologies Agilent Technologies will provide warranty services as described in Table 45.
Safety Information Safety Information The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
Safety Information rendering first aid and resuscitation, is present. Do not replace components with power cable connected. Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard. Do not install substitute parts or make any unauthorized modification to the instrument.
Safety Information WAR NI N G A warning alerts you to situations that could cause physical injury or damage to the equipment. Do not proceed beyond a warning until you have fully understood and met the indicated conditions. CA UT IO N A caution alerts you to situations that could cause a possible loss of data. Do not proceed beyond a caution until you have fully understood and met the indicated conditions.
Lithium Batteries Information Lithium Batteries Information WAR NI N G Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended by the equipment manufacturer. Lithium batteries may not be disposed-off into the domestic waste. Transportation of discharged Lithium batteries through carriers regulated by IATA/ICAO, ADR, RID, IMDG is not allowed.
Radio Interference Radio Interference Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations. Test and Measurement If test and measurement equipment is operated with equipment unscreened cables and/or used for measurements on open set-ups, the user has to assure that under operating conditions the radio interference limits are still met within the premises.
Solvent Information Solvent Information Observe the following recommendations on the use of solvents. Flow Cell Avoid the use of alkaline solutions (pH > 11) which can attack quartz and thus impair the optical properties of the flow cell. Solvents Always filter solvents, small particles can permanently block the capillaries. Avoid the use of the following steel-corrosive solvents: • Solutions of alkali halides and their respective acids (for example, lithium iodide, potassium chloride, and so on).
Agilent Technologies on Internet Agilent Technologies on Internet For the latest information on products and services visit our worldwide web site on the Internet at: http://www.agilent.com/go/chem Select “Products” - “Chemical Analysis” It will also provide the latest firmware of the Agilent 1100 series modules for download.
Index A accessory kit, 18 accessory kit, degasser, 19 active inlet valve, 103, 178 active inlet valve drive, 210 air flow, 141 algae growth, 35 alternative seal material, 39 ambient non-operating temperature, 16 ambient operating temperature, 16 analog output, 255 APG remote connector, 23, 220 ASIC - application-specific integrated circuit, 210 assembling the main cover, 155 AUTO mode, 203 AUX output, 23 compensation sensor short, 56 composition precision, 254 composition range, 254 compressibility compen
Index hydraulic path, 166 hydraulic system, 254 pistons, 102, 115 power supply, 147 pump drive, 143 pump seals, 102, 112 purge valve, 102, 108 purge valve frit, 102, 108 status light pipe, 154 wash seals, 102, 116 F fan, 141 fan drive, 211 fan failed, 57 fan out, 220 features GLP, 255 instrument layout, 206 safety and maintenace, 255 firmware description, 214 main system, 214 resident system, 214 firmware update with control module, 246 flow connections, 25 flow precision, 254 flow range, 254 foam, 156 fo
Index bottle head assembly, 165 control module, 162 cover, 167 damaged, 17 foam, 169 hydraulic path, 166 leak handling, 171 light pipes, 170 main assemblies, 160 missing, 17 outlet ball valve, 176 pump head, 172 pump head with seal wash, 174 pump housing and main assemblies, 161, 162 sheet metal kit, 168 solvent cabinet, 164 parts identification cables - LAN cables, 195 performance specification, 254 pH range, 254 physical specifications, 16 piston, 32, 115, 200 piston chamber, 199 plateaus, leak test, 93
Index wrench, 14 mm, 18 status indicator, 44, 45 status lamp, 45 status light pipe, 154, 170 stay resident settings, 226 stroke length, 79 stroke volume, 200, 203 sychronization lost, 50 syringe, 19 syringe adapter, 19, 28 Z zero solvent counter, 60 T temperature limit exceeded, 73 temperature out of range, 72 temperature sensor, 52 tests on control module, 251 timeout, 48 top cover, 155, 156 U unknown error 2055, 60 unpacking the pump, 17 V vacuum degasser, 17, 25, 32, 198 variable reluctance motor, 20
sa In This Book This manual contains technical reference information about the Agilent 1100 Series quaternary pump. The manual describes the following: • installation, • optimizing performance, • diagnostics and troubleshooting, • repairing, • parts and materials, • theory of operation, and • screens of local control module.