Technical data

134 Agilent U1251B/U1252B User’s and Service Guide
6 Performance Tests and Calibration
Ta bl e 6 - 4 Adjustment table
Step Test Function Cal Range Input Cal Item
U1251B U1252B
1 Turn the rotary switch to the
position
5V 0.3 V,1 kHz 0.3000 V 0.3000 V
3 V, 1 kHz 3.0000 V 3.0000 V
3 V, 10 kHz 3.0000 V 3.0000 V
50V 3 V, 1 kHz 03.000 V 03.000 V
30 V, 1 kHz 30.000 V 30.000 V
30 V, 10 kHz 3.0000 V 30.000 V
500V 30 V,1 kHz 030.00 V 030.00 V
300 V,1 kHz 300.00 V 300.00 V
300 V, 10 kHz 3.0000 V 300.00 V
1000V 30 V, 1 kHz 0030.0 V 0030.0 V
300 V, 1 kHz 0300.0 V 0300.0 V
300 V, 10 kHz 3.0000 V 0300.0 V
2 Turn the rotary switch to
position (for model
U1252B), to
position
(for model U1251B)
Short Dual Banana Plug
with copper wire
short between 2
terminals
SHort SHort
5V 3 V 3.0000 V 3.0000 V
50V 30 V 30.000 V 30.000 V
500V 300 V 300.00 V 300.00 V
1000V 1000 V 1000.0 V 1000.0 V