Technical data
72 Agilent U1251B/U1252B User’s and Service Guide
4 Changing The Default Setting
Menu item
Available setting options Default factory
setting
Display
Description
Display Description
Print
Print
On, OFF Enables auto send of data to the PC continuously
when set to On
OFF
rESEt
Reset
dEFAU Enables the reset of factory settings by pressing
and holding for more than 1 second
dEFAU
Te mp er a tu re
(4) d-CF Sets the temperature measurement to °C but
pressing swaps the display to °F
d-C
d-F Sets the temperature measurement to °F
d-FC Sets the temperature measurement to °F but
pressing swaps display to °C
d-C Sets the temperature measurement to °C
bAtt
Battery voltage
7.2 V, 8.4 V Selects battery voltage to either 7.2 V or 8.4 V 7.2 V
FiLtE
DC filter
On, OFF Enables DC filter when set to On OFF
NOTE
1 This is the first display once the meter enters the Setup mode.
2 For d-LoG and rEF menu items, press to select the digit to be adjusted.
3 J-type thermocouple applies to U1252B.
4 To view the menu item, press for more than 1 second.