Technical data

Calibration Procedures 1
Performance Verification Tests
U1270 Series Service Guide 27
Performance Verification Tests
Use the performance verification tests to verify the
measurement performance of the instrument. The
performance verification tests use the instrument's
specifications listed in the U1271A/U1272A User's Guide and
the U1273A/U1273AX User’s Guide (available for download at
www.agilent.com/find/hhTechLib).
The performance verification tests are recommended as
acceptance tests when you first receive the instrument. The
acceptance test results should be compared against the
1 year test limits. After acceptance, you should repeat the
performance verification tests at every calibration interval.
If the instrument fails the performance verification tests,
adjustment or repair is required.
Ensure that you have read the “Test Considerationson page 26 before
running the performance verification tests.