Technical data

1 Calibration Procedures
Performance Verification Tests
32 U1270 Series Service Guide
AC+DCV Turn the rotary
switch to the position, and
press the key twice.
1000 V 1000 V, 45 Hz - ±9.5 V
1000 V, 65 Hz - ±9.5 V
1000 V, 1 kHz - ±14.0 V
1000 V, 5 kHz - ±19.0 V
8
DCmV
[2]
Turn the rotary
switch to the / position.
30 mV 30 mV - ±0.035 mV
–30 mV - ±0.035 mV
300 mV 300 mV ±0.20 mV ±0.20 mV
–300 mV ±0.20 mV ±0.20 mV
[2] The accuracy is specified after the Null function is used to subtract the thermal effect (by shorting the test leads) before
measuring the signal.
9
AC+DCmV Turn th e
rotary switch to the
position, and press the
key twice.
30 mV 30 mV, 20 Hz - ±0.285 mV
30 mV, 45 Hz - ±0.250 mV
30 mV, 65 Hz - ±0.250 mV
30 mV, 1 kHz - ±0.285 mV
30 mV, 5 kHz - ±0.375 mV
30 mV, 20 kHz - ±0.390 mV
30 mV, 100 kHz - ±1.140 mV
Table 1 - 2 Performance verification tests (continued)