Check the Fuse
U1273A/U1273AX Quick Start Guide 15
Check the Fuse
Follow the instructions below for a quick check on the fuses
(Fuse 1 and Fuse 2) of your multimeter.
Fuse
Part
number
Fuse rating
Displayed readings
Fuse healthy Replace fuse
1 2110-1400 440 mA/1000 V 102 Ω OL
2 2110-1402 11 A/1000 V 0.05 Ω OL
Smart
Smart
To check Fuse 1 To check Fuse 2