To disable OVP
1 Press on the “Over Voltage” button to disable the OVP.
The OVP annunciator will eventually turn off.
To clear the over voltage condition
1 OVP trip can only occur when the output is enabled.
When the OVP condition occurs:
The output is disabled and the OFF annunciator will be
turned on.
If the keylock feature was enabled, it will be disabled.
The LOCK annuciator will be turned off.
The OVP annunciator will fast- blink continuously.
The display will show “TRIP”.
2 To clear the OVP trip conditon, press on the “Over
Voltage” button again.
NOTE
If OVP trip persists, decrease the voltage limit settings to clear the
tripping.