Technical data

Service Guide 5
U8001A/U8002A User’s and Service Guide 65
Test Consideration
For optimum performance, all procedures should comply
with the following recommendations:
Assure that the calibration ambient temperature is stable
and between 20 °C and 30 °C. Ideally the calibration
should be performed at 25 °C ±1 °C.
Assure ambient relative humidity is less than 80%.
Allow one hour warm up period by power on the power
supply. During warm up period, set the power supply to
maximum programmable voltage and current. Then, enable
the output.
Keep the test setup connection cables as short as possible.
Resistive Load (R
L
) 10 Ω/90 W (for U8001A)
6 Ω/150 W (for U8002A)
Measures ripple and
noise
PV
Current Monitoring
Resistor (Shunt)
(R
M)
0.01 Ω ± 0.1%
TCR less than 20 ppm/ºC
ISOTEK Co.
Model A-H
Current monitoring
resistor
PV
C
* PV = Performance Verification
C = Calibration
Table 5 - 7 List of Equipment
Equipment Requirement(s) Recommended Model Purpose Used