Current Monitoring Resistor
To eliminate output current measurement error caused by 
the voltage drops in the leads and connections, connect the 
current monitoring resistor (R
M
) between the (–) output 
terminal and the load as a four- terminal device. Connect the 
current monitoring leads inside the load- lead connections 
directly at the monitoring points on the resistor element.
General Measurement Techniques
To achieve best results when measuring load regulation, peak 
to peak voltage, and transient response time of the power 
supply, measuring devices must be connected through the 
hole in the neck of the binding post at (A) while the load 
resistor is plugged into the front of the output terminals at 
(B). A measurement made across the load includes the 
impedance of the leads to the load. The impedance of the 
load leads can easily be several orders of the magnitude 
greater than the power supply impedance and thus 
invalidate the measurement. To avoid mutual coupling 
effects, each measuring device must be connected directly to 
the output terminals by separate pairs of leads.
Figure 5-8 General Measurement