CC Load Effect (Load Regulation)
This test measures the immediate change in output current
resulting from a change in the load from full rated output
voltage to short circuit.
Procedures:
1 Power off the power supply and connect the output to be
tested as shown in Figure 5- 7(B) with the DMM
connected across the 0.01
Ω current monitoring resistor
(R
M
).
2 Power on the power supply.
3 When the power supply is in limit mode, program the
output voltage to maximum programmable value and
output current to full rated value, i.e. 3 A for U8001A and
5A for U8002A.
4 Enable the output.
5 Operate the electronic load in constant voltage mode and
set its voltage to 30 V. Ensure that the power supply is in
CC mode. If not, adjust the electronic load so that the
voltage drops slightly until the power supply is in CC
mode.
6 Record the current reading by dividing the voltage reading
on the DMM by the resistance of the current monitoring
resistor.
7 Within a few seconds after step 6, operate the electronic
load in short mode and record the current reading by
dividing the voltage reading on the DMM by the resistance
of the current monitoring resistor.
8 Compare the reading obtained in steps 6 and 7. The
difference should be within the limit of:
• U8001A: 2.6 mA
• U8002A: 3 mA