Technical data

80 U8001A/U8002A User’s and Service Guide
5Service Guide
9 Compare the reading obtained in steps 7 and 8. The
difference should be within the limit of:
U8001A: 2.6 mA
U8002A: 3 mA
Line Voltage Low Line Voltage Limit, V
ac
High Line Voltage Limit, V
ac
100 90 110
115 104 127
230 207 253