U8001A/U8002A User’s and Service Guide 81
CC Noise
CC noise is specified as the RMS output current in a 
frequency range 20 Hz to 20 MHz with the power supply in 
constant current operation.
Procedures:
1 Power off the power supply and connect the output to be 
tested as shown in  Figure 5- 7(D) to an AC/DC converter 
and load resistor (10
 Ω for U8001A and 6 Ω for U8002A).
2 Power on the power supply.
3 When the display is in the limit mode, program the 
output voltage to maximum programmable value and 
output current to full rated value, i.e. 3 A for U8001A and 
5 A for U8002A. 
4 Enable the output. 
5 Make sure that the power supply is in CC mode. If not, 
adjust the resistive load so that the output voltage drops 
slightly until the power supply is in CC mode.
6 Configure the differential amplifier as follow:
• Set to AC mode (positive and negative) to removes the 
DC component
• Set to differential mode
• Set gain to x10
• Set attenuation to 1
• Set low pass filter to 20 MHz bandwidth limit to filter 
out input signal which contains higher frequency
• Set to zero precision voltage generator
• Set input impedance to 1 M
Ω
7 Configure the RMS voltmeter as follow:
• Set high pass filter to10 Hz
• AC coupling 
8 Divide the reading of the RMS voltmeter by 10 and then 
multiply the result by 5 to get the RMS noise 
measurement. The result should not exceed the RMS 
limits of 3 mA
RMS
.