Technical data

Service Guide 5
U8001A/U8002A User’s and Service Guide 83
4 High Voltage Calibration Point.
Display will show "C" blinking.
The M3 annunciator will be turned on.
Use the knob to enter the reading obtained from the
DVM.
Press the "Memory" button to save the changes voltage
calibration is done. You will be able to perform this if
the power supply is in CV mode.
5 Voltage Calibration
Display will show "donE” for a few seconds and the
power supply will enter the current calibration mode.
NOTE
If you do not wish to continue with the current calibration, you may now
turn off the power supply to exit the calibration mode.