OPERATING MANUAL 300 WATT ELECTRONIC LOAD MODULE Agilent Model 60502B FOR MODULES WITH SERIAL NUMBERS: 3118A-00101 AND ABOVE Agilent Part No. 60502-90008 Microfiche No. 60502-90009 Printed in U.S.A.
DECLARATION OF CONFORMITY according to ISO/IEC Guide 22 and EN 45014 Manufacturer’s Name: Agilent Technologies Manufacturer’s Address: New Jersey Division 140 Green Pond Road Rockaway, NJ 07866 U.S.A.
00-Watt Module About This Manual This manual provides information for the Agilent 60502B 300-Watt Electronic Load Module. It is designed as a supplement to the Agilent 6050A/6051A Multiple Input Mainframe Electronic Load Operating Manual (part number 0605090001). Four tables provide the following module-specific information: Table 60502-1 lists both the specifications and supplemental characteristics of the module.
Table 60502-1. Specification and Supplemental Characteristics SPECIFICATIONS DC Input Rating: Current: 0 to 60 A Voltage: 3 V to 60 V (minimum dc operation from 0 to 2 V for 0 to 60 A) Power: 300 W at 40 °C (derated to 225 W at 55 °C) A. OPERATING CHARACTERISTICS B. DERATED CURRENT DETAIL Constant Current Mode: Ranges: Accuracy: Resolution: Regulation: Temperature Coefficient: 0 to 6 A; and 0 to 60 A (after 30 second wait): ± 0.1% ± 75 mA (both ranges) 1.
Table 60502-1 Specifications and Supplemental Characteristics (continued) Transient Operation: Continuous Mode Frequency Range: Frequency Resolution: Frequency Accuracy: Duty Cycle Range: Duty Cycle Resolution: Duty Cycle Accuracy: 0.25 Hz to 10 kHz 4% 3% 3% to 97% (0.
Table 60502-1 Specifications and Supplemental Characteristics (continued) External Analog Programming 0 to 10 V (dc or ac): Bandwidth: Accuracy: Temperature Coefficient: 10 kHz (3 db frequency) ± 4.5% ± 75 mA (0 to 6 A range) ± 4.5% ± 250 mA (0 to 60 A range) ± 0.8% ± 200 mV (0 to 60 V range) 100 ppm/°C ± 6 mA/°C (current ranges) 100 ppm/°C ± 1 mV/°C (voltage range) External Current Monitor (0 to 10 V): Accuracy: Temperature Coefficient: ± 0.
Table 60502-1 Specifications and Supplemental Characteristics (continued) Current Slew Rate:* Rate # 1 2 3 4 5 6 7 8 9 10 11 12 60 A Range Step 6 A Range Step Transition Time 1 A/ms 0.1 A/s 8.0 ms 2.5 A/ms 0.25 A/s 3.2 ms 5 A/ms 0.5 A/s 1.6 ms 10 A/ms 1 A/ms 800 µs 25 A/ms 2.5 A/ms 320 µs 50 A/ms 5 A/ms 160 µs 10 A/ms 0.1 A/µs 80 µs 25 A/ms 0.25 A/µs 32 µs 50 A/ms 0.5 A/µs 16 µs 1 A/µs 0.1 A/µs 12 µs 2.5 A/µs 0.25 A/µs 12 µs 5 A/µs 0.5 A/µs 12 µs *AC performance specified from 3 to 60 V.
Table 60502-1 Specifications and Supplemental Characteristics (continued) Source Turn-On Current Overshoot: Less than 10% of final value (in CC and CR modes when connected to power supplies with voltage rise times of greater than 500µs). Programmable Short Circuit: 0.033 Ω (0.002 Ω typical) Programmable Open Circuit: 20 kΩ (typical) Drift Stability (over an 8 hour interval): Current: Voltage: ± 0.03% ± 10 mA ± 0.
Function Transient Operation Set Frequency Set Duty Cycle *Set Pulse Width Trigger Operation *Set Trigger Period Current Protection *Set Current Level *Set Delay Time Table 60502-2. Programming Ranges (continued) Front Panel Front Panel HPSL Command Key Display (Short Form) (shift) FREQ value DCYCLE value "TRAN:FREQ value" "TRAN:DCYC value" "TRAN:TWID value" 0.25 Hz to 10 kHz 3-97% (0.25 Hz-1 kHz) 6-94% (1 kHz-10 kHz) 0.00005 to 4 s "TRIG:TIM value" 0.
Ranges and Calibration Points High Current Range High Current Offset Low Current Range Low Current Offset Voltage Range Voltage Hi point Voltage Lo point Low Resistance Range Low Resistance Hi point Low Resistance Lo point Middle Resistance Range Middle Resistance Hi point Middle Resistance Lo point High Resistance Range High Resistance Hi point High Resistance Lo point Table 60502-4. Calibration Information Variables Variables Power Supply Value Settings Hi_curr_rng 60 5 V/61 A Hi_curr_offset 0.
Agilent Sales and Support Office For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir You can also contact one of the following centers and ask for a test and measurement sales representative. United States: Agilent Technologies Test and Measurement Call Center P.O.