Chapter 1, Description of Keys
Keys
Ref set
Press and release the Shift key, then the Incrementรท10 key to access the reference
set function. This function is used to enter or remove a measurement reference for
relative AF and RF measurements.
Programming Ref set
The GPIB command :REFerence is used to select this data function
programmatically.
To Turn Measurement Reference Points ON and OFF. Use the
:REFerence:STATe <boolean> commands to turn measurement reference points
ON and OFF. The <boolean> parameter can be a 1 or ON to turn measurement
reference points on, and a 0 or OFF to turn measurement reference points off.
Syntax
:REFerence:STATe <ON> or <1>
:REFerence:STATe <OFF> or <0>
Example
OUTPUT 714;"MEAS:RFR:POW:REF:STAT ON"
This turns the measurement reference point for the TX power measurement
ON.
To Query the State of Measurement Reference Points. Use the
:REFerence:STATe? commands to query the state of a measurement reference
point. The query returns a 1 if a measurement reference points is ON, and a 0 if a
measurement reference points is OFF.
Syntax
:REFerence:STATe?
Example
OUTPUT 714;"MEAS:RFR:POW:REF:STAT?"
ENTER 714;Meter_on_off ! returns a 1 (ON) or 0 (OFF)
This queries the state of the measurement reference point for the TX power
measurement.