Operating instructions

Select by pressing key ... Enter °C s. Sec. 5.4
FUNCTION
1: 1 8 0.0 TK
6.3 Measuring point scans and data output
Measuring point scans can be used to acquire and then evaluate data from all
active measuring points at particular times (for the purposes e.g. of averaging,
outputting to the interface, or saving to memory). A measuring point scan can
be initiated at any time-of-day either manually or automatically by a cycle. If a
peripheral device (e.g. printer) is connected (see Manual, Section 5.2), each
measured value is output to it via the interface in the output format set (see
Section 6.3.3) (printout, see Section 6.6.1). In time-critical applications it may
be preferable to activate continuous measuring point scanning (see Manual,
Section 6.5.1.3); scanning all measuring points in continuous mode takes only
half as long as in semi-continuous mode; (see Section 6.1, default setting ´con-
tinuous´ off).
6.3.1 Once-only measuring point scan (see Manual, Section 6.5.1.1)
A once-only manual measuring point scan for acquiring instantaneous measu-
red values from all active measuring points can only be initiated in the MEASU-
RED VALUE function by pressing the MANU/PRINT key. The ´START´ arrow
lights up briefly; and if the interface is connected, the ´COM´ arrow should also
appear. With the “S” option the measured values are saved and the ´MEM´ ar-
row lights up. Each time the key is pressed again after this the measured valu-
es are processed accordingly. If the actual time-of-day is to be indicated as
measuring time, this needs to be entered first; (see Section 9.1).
Once-only measuring point scan
The MEASURED VALUE function by :
6.3.2 Cyclic measuring point scan (see also Manual, Section 6.5.1.2)
For cyclic measuring point scans and outputs the output cycle can be set in the
CYCLE function ´CY´.
CYCLE function ´CY´
Select by pressing the FUNCTION key ...
Enter the time in the 6-character format hh:mm:ss (see Section 5.4).
Example
: print cycle 30 minutes
Delete cycle by pressing the keys: ,
PROG
CLEAR
This terminates any cyclic scan currently still running.
MANU/
PRINT
FUNCTION
0 0:3 0:0 0.ZY
ALMEMO
2390-5 29
Measuring operations