User's Guide
14 
  If you only intend to use the blood glucose meter and do not want to calibrate 
the CGMS, press "Record" after the test result is displayed. Your blood glucose 
reading will be recorded to history, but not used to calibrate the real
-time 
glucose monitoring system. 
  You can also choose to press  “Manual Input”  when  “Please insert blood 
glucose test strip”appears. This allows you to input a reference blood glucose 
value from a different blood glucose meter. Press“Calibrate”  to complete the 
calibration. 
  If you do not wish to calibrate the CGMS, you can click  “Record”  to enter a 
blood glucose reading into history. The system will only record the current blood 
glucose value and measurement time, but it will not calibrate the CGMS。 
  The blood glucose test interface will appear any time you insert a blood glucose 
test strip, even when the PDA is active and on the home screen. 
  It is recommended to perform a calibration as soon as possible after the sensor 
is initialized to confirm the performance of the sensor. 
  If you doubt that the glucose level displayed by the CGMS is correct, you can use 
the calibration function to confirm the performance. However, do not calibrate 
the system frequently within a short amount of time. 
Removing a Sensor 
Remove an old sensor when the PDA indicates that the sensor has expired or if you 
feel any site irritation or discomfort. 
  Carefully pull the tape that holds the sensor to the skin from the edge and slowly 
uncover it until the entire sensor is removed. You can use warm soapy water to 