User's Guide
15 
remove any remaining sticky residue. 
  Pry the two locking arms on  one side of the  sensor base, push the transmitter 
away  from the  sensor  base,  and  then  remove  thetransmitter  so  that  it  can  be 
used again. 
  The sensor is meant for one-time use only. Dispose of the used sensor according 
to your local regulations. 
【PDA Settings】 
Target Glucose Range 
  You can change the thresholds for high glucose alert and low sensor 
glucose alert. These settings change the normal range of sensor glucose 
concentrations displayed on the home page and when high/low glucose alerts 
are triggered. 
  Before setting the target glucose range, please make sure that the 
transmitter is connected to the sensor base, and that the PDA and 
transmitter have a good wireless connection. 
  Press "Settings" on the PDA’s home screen to enter the Settings Menu.; 
  Press“High Glucose Threshold”to enter the desired sensor blood 
glucose threshold. The threshold range is 8-25mmol/L (default value 
is 12mmol/L). 
  Press "Low Glucose Threshold" to enter the desired low sensor glucose 
threshold.  The  threshold  range  is  2.2-5mmol/L  (default  value  is 
3.5mmol/L). 