User's Guide
25 
the range printed on the test strip package (typically CTRL1) the 
device isworking properly. 
  After the test is finished, eject the test strip. If the control 
solution results are outside of the reference range: 
  Confirm you are matching the correct range. Control Solution 1 
results should be matched to the CTRL1 range printed on the test 
strip vial (or foil pouch). 
  Check the expiration date of the test strip and control solution. 
Make sure that the packages have not been opened for more than 6 
months. Discard any expired test strips and control solution. 
  Confirm that you are testing within the correct temperature range 
(15-30 °C). 
  Make sure that the test strip vial and control solution bottle have 
been tightly closed. 
  Make sure that you are using the correct brand of control solution. 
  Make sure you are following the user guide instructions properly. 
After checking all of the conditions above, repeat the quality control 
test with a new test strip. If the quality control test results are still 
outside of the range printed on the test vial (or foil pouch), there may 
be a problem with your meter. Please seek help and contact your dealer. 
Control Solution 1 is sufficient for most self testing needs. If you 
think your meter or strips may not be working correctly, you may also want 
to do a level 2 test. The ranges for both (CTRL 1 and CTRL 2) are displayed 
on the test strip vial (or on the foil pouch). Simply repeat the test 
procedure, using Control Solution 2. For confirmation of results, Control 
Solution 1 tests should fall within the CTRL1 range, and ControlSolution 
2 tests should fall within the CTRL2 range.。 